Depth of immersion (multi parameter probe MPP 9x0 IDS)
Multi 3630 IDS
70
ba77170e03
10/2017
10 Depth of immersion (multi parameter probe
MPP 9x0 IDS)
10.1 General information
With a multi parameter probe of the series MPP 9x0 IDS and the corresponding
IDS sensor of the XXX-P series you can add the parameter depth of immersion
(DPT) to the display of your Multi 3630 IDS meter. The depth of immersion is
integrated as a secondary parameter in the display of all main measured
parameters.
Fig. 10-1 Example: Depth of immersion as main and secondary measured parameter
(MPP 9x0 IDS with a pH-IDS sensor of the XXX-P series)
With the maximum equipment (MPP 930 with 3 IDS sensors of the XXX-P
series) you can at the same time record three main parameters (pH, O2,
ϰ
) and
two secondary parameters (temperature and depth of immersion).
Special features of
the multi parameter
probes MPP 9x0 IDS
If the MPP 930 is equipped with the maximum number of IDS sensors of the
XXX-P series (3), the setting menu for the parameter depth of immersion is
only available from the single display of the parameter (
<F2>
+
<
><
>
,
If a multi parameter probe MPP 9x0 IDS is connected to the meter, no other
sensor can be operated while connected directly to the meter. Display:
Open plug connections of the multi parameter probe MPP 9x0 IDS can
cause damage if they come into contact with water. Therefore, they always
have to be closed with a blind plug (BPI/DS 900). Display:
Sensor input
open!
As soon as an IDS sensor of the XXX-P series is connected to the multi
parameter probe MPP 9x0 IDS, the parameter depth of immersion is inte-
grated as a secondary parameter into the measured value display of the
main parameter.
Depth values are displayed from a depth of immersion of 0.5 m.
Depending on the equipping of the MPP 9x0 IDS, the following value is dis-
played as the depth of immersion:
1 Depth of immersion as the main measured value
2 Depth of immersion as a secondary measured value
1
2
Содержание WTW OxiTop-IDS Multi 3630 IDS
Страница 1: ...OPERATING MANUAL ba77170e03 10 2017 Multi 3630 IDS DIGITAL METER FOR IDS SENSORS 3630 IDS...
Страница 2: ...Multi 3630 IDS 2 ba77170e03 10 2017 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...
Страница 15: ...Multi 3630 IDS Commissioning ba77170e03 10 2017 15 3 3 4 Setting the date and time 3 See section 4 5 5...
Страница 117: ......