Oxi 3205
Technical data
35
ba75797e05
12/2017
7
Technical data
7.1
General data
Dimensions
approx. 180 x 80 x 55 mm
Weight
approx. 0.4 kg
Mechanical structure
Type of protection
IP 67
Electrical safety
Protective class
III
Test certificates
CE
Ambient
conditions
Storage
- 25 °C ... + 65 °C
Operation
-10 °C ... + 55 °C
Allowable relative
humidity
Annual mean: < 75 %
30 days/year: 95 %
Other days: 85 %
Power
supply
Batteries
4 x 1.5 V alkali-manganese batteries, type
AA
Rechargeable
batteries
4 x 1,2 V NiMH rechargeable batteries,
type AA (no charging function)
Operational life
up to 800 h without / 100 h with illumination
Guidelines
and norms used
EMC
EC directive 2004/108/EC
EN 61326-1
EN 61000-3-2
EN 61000-3-3
FCC Class A
Meter safety
EC directive 2006/95/EC
EN 61010-1
IP protection class
EN 60529