NitraVis 70x IQ TS
Maintenance and cleaning
35
ba75952e03
05/2017
5
Maintenance and cleaning
5.1
Maintenance
The UV-VIS- NitraVis 70x IQ TS sensor operates maintenance-free.
5.2
Sensor cleaning
5.2.1 Cleaning agents and accessories
Cleaning agents To clean the sensor, use the following cleaning agents only:
* For the final cleaning prior to a
Sensor check
or
Zero adjustment
, only use
cleaning agents with purity grade "for analysis". Finally, thoroughly rinse the
sensor with ultrapure water.
Accessories
Sponge
Plastic-bristled brushes of different sizes.
Interdental brushes are suitable for the cleaning of the compressed air chan
-
nels.
Flocked cleaning cards (see accessories)
Lens tissues
Wash bottle
NOTE
Never use any pressure washer, steam blaster, abrasives or metal brushes to
clean the sensor. Risk of damaging the sensor.
Contamination
Cleaning agents
Water-soluble substances
– Primary cleaning: Tapwater, warm if
possible
(30 - 50 °C)
– Precision cleaning: Household wash
-
ing-up ultrapure water.
Lime deposits
5 % hydrochloric acid *
Grease deposits
Mixture of isopropanol and water (approx.
70 % isopropanol) *
Содержание WTW NitraVis 700 Series