Measurement characteristics
Measuring principle Spectrophotometric Absorption measurement; integrated microprocessor elec
-
tronics, shielded 2-wire connection for power and data transmission.
Light source
Detector
Wavelengths range 200 - 720 nm
Measuring gap
(optical layer
thickness)
TSS measurement,
5 mm gap width
Nitrate
measurement,
5 mm gap width
Carbon
measurement,
5 mm gap width
Lamp type
Xenon flashlamp
Photo diodes
256
NiCaVis 705 IQ TS
5 mm
Measuring location
Measuring mode
Measuring range
Resolution
Outlet
TSS
0.0 ... 900.0 mg/L
0.1 mg/L
Measuring location Measuring mode
Measuring range
Resolution
Outlet
NO3
0.0 ... 250.0 mg/L
0.1 mg/L
NO3-N
0.00 ... 50.00 mg/L 0.01 mg/L
Measuring location
Measuring mode
Measuring range
Resolution
Outlet
COD spectr., total
0.0 ... 800.0 mg/L 0.1 mg/L
COD spectr., dissolv
0.0 ... 800.0 mg/L 0.1 mg/L
TOC spectr., correl.
0.0 ... 500.0 mg/L 0.1 mg/L
DOC spectr., correl.
0.0 ... 500.0 mg/L 0.1 mg/L
BOD spectr., correl.
0.0 ... 500.0 mg/L 0.1 mg/L
SAC 254, total
0.0 ... 600.0 1/m
0.1 1/m
SAC 254, dissolved
0.0 ... 600.0 1/m
0.1 1/m
UVT 254, total
*
* The UVT-254 value is standardized to 10 mm gap width.
0.0 ... 100.0 %
0.1 %
UVT 254, dissolved
*
0.0 ... 100.0 %
0.1 %