pH value
Multi 3320
28
ba77053e04
07/2018
5.2.4
Manual calibration (
ConCal
)
Make sure that in the sensor menu,
Buffer
menu, the ConCal buffer set is
correctly selected (see section 10.1.1 S
Use any one to five buffer solutions in ascending or descending order.
21. Wait for the measurement with stability control to be completed or termi-
nate the stability control and take over the calibration value with
<ENTER>
.
The calibration display for the next buffer appears (voltage display).
22. If necessary, use
<M>
to finish calibration or
Continue calibrating using the next buffer with
<ENTER>
.
Calibration is automatically completed after the last buffer of a buf-
fer set has been measured. Then the calibration record is dis-
played.
The calibration line is determined by linear regression.
If single-point calibration was set in the menu, the calibration proce-
dure is automatically finished with the measurement of buffer solu-
tion 1 and the calibration record is displayed.
1.
Connect the pH electrode to the meter.
The pH measuring window is displayed.
2.
When measuring without temperature sensor:
– Temper the buffer solution, or measure the current temperature.
– Enter the temperature value in the menu.
3.
In the measured value display, select the measured parameter pH or mV
with
<M>
.
4.
Start the calibration with
<CAL>
.
The calibration display appears (voltage display).
pH
180.0
24.8
°C
AutoCal TEC
4.010
15.02.2014
14:15
AR
Buffer
mV
Содержание WTW Multi 3320
Страница 1: ...OPERATING MANUAL ba77053d04 07 2018 Multi 3320 METER FOR 2 SENSORS PH ORP D O COND 3320 ...
Страница 8: ...Overview Multi 3320 8 ba77053e04 07 2018 ...
Страница 102: ...Glossary Multi 3320 102 ba77053e04 07 2018 ...
Страница 104: ...Index Multi 3320 104 ba77053e04 07 2018 Two point calibration ISE 42 pH 26 30 Z Zero point of pH electrode 25 ...
Страница 105: ......