MIQ/IDS
Dissolved oxygen
33
ba77277d01
01/2024
Canceling the cali
-
bration
As long as the determination of the calibration data has not yet been started
(Step 2:, page 31), you can quit the calibration routine with
<M>
or
<ESC>
.
You can cancel the ongoing calibration data determination as follows:
After the calibration is canceled, the sensor works again with the calibration data
that was used before the canceled calibration.
6.2.4 Calibration result
Possible calibration
results
The calibration data are evaluated by the IQ S
ENSOR
N
ET
. A calibration proce-
dure can have the following results:
Calibration history The calibration history contains the data of the active calibration and the chrono
-
logical list with the last calibrations.
The calibration history provides the following information:
1.
Open the setting menu of the sensor (see section 6.3).
2.
In the
Calibration
menu, select the setting
abort
and then exit the setting
table with
Save and quit
.
Display
Explanation
Measured value display Sensor was successfully calibrated.
The calibration data can be viewed in the calibration
history.
"----"
The sensor could not be calibrated.
The sensor is blocked for further measurement.
Notes on possible causes are given in the log book
of the sensor.
Date
Date of the calibration
Rel.slope
Relative slope (non-dimensional)
Evaluation of the calibration
o.k.
: Successful calibration.
The new calibration data are taken over for measurement.
Error
: Calibration unsuccessful.
Invalid calibration data were discarded. Measurement was con
-
tinued with the last valid values.
default
: Default value (delivery status)