DIQ/S 281
Installation
19
ba77244d01
08/2020
figure 3-4
SENSORNET connection
3.6
On-site mounting of the DIQ/S 28X
3.6.1
General information
The universal transmitter DIQ/S 28X is designed for stationary installation. With
the aid of mounting accessories, the setup can be adjusted to meet various
requirements.
Components installed outside must always be protected by a sun shield against
the effects of the weather (snow, ice and direct solar radiation). Otherwise, mal-
5
Connect the cable ends to the terminal strip. When doing so, look out
for the designations of the terminals (white & brown / shield / green).
6
Tighten the cap nut (pos. 3 in figure 3-1).
7
Close the enclosure.
The complete assignment of the terminal strip is shown in
section 3.12.
For further instructions on the mounting of the sensor at the applica-
tion location, please refer to the relevant manual (immersion depth,
etc.).
Sensor
X3
X2
X1
SENSORNET
RED
RO
T
SHIELD
SCHIRM
GREEN
GRÜN
red
green
b
l
ack