150
Technical Specifications
1.3
1.3.1
Titrator TitroLine
®
7800
Translation of the legally binding German version
(Release: 16. August 2018)
EMC compatibility according to the Council Directive: 2014/30/EU;
applied harmonized standards: EN 61326-1: 2013
Low-voltage directive according to the Council Directive 2014/35/EU;
Testing basis EN 61 010-1: 2010 for laboratory equipment
RoHS Council Directive 2011/65/EU
FCC Part 15B and ICES 003
Country of origin
: Germany, Made in Germany
The following solvents/titration reagents are allowed to be used:
All common titration solutions.
As reagent water and all non-aggressive non-organic and organic fluids are allowed.
If using
combustible fluids fire please adhere to the Guidelines for Explosion Protection and Prevention of the
chemical industry.
For fluids with higher viscosity (
5 mm
2
/s), lower boiling point or affinity to outgas, the filling and dosage
speed can be adjusted.
Fluids with viscosity over 20mm
2
/s cannot be dosed.
To ensure maximum accuracy of the readings we recommend to allow some reasonable time for the
TitroLine
®
7800
to “warm up”.
Measuring input 1
(analog):
pH/mV-input with 24 bit transducer for high-precision readings.
Electrode socket according to DIN 19 262 or additional with BNC socket insert (Z 860).
Reference electrode 1 x 4 mm socket.
Adjustable damping settings of the pH/mV measuring signal.
RFID receiver for SI Analytics
®
ID electrodes.
Measurement range
Over range
Display resolution
Measurement accuracy*
without sensor probe
Input
resistance
pH
pH
- 3.0 ... 18.00
- 3.
1 … 18.00
0.001
0.002
1 Digit
> 1
10
13
mV
U
mV
- 2000 ... 2000
- 2020 ... 2020
0.1
0.10
1 Digit
> 1
10
13
Measuring input
(Pt 1000):
Temperature sensor - connector for a resistance thermometer Pt 1000
and NTC 30 kOhm. Connection: 2 x 4 mm sockets.
Measurement
range T [°C]
Display resolution
Measurement accuracy* without
sensor probe
Pt 1000
- 75 ... 195
0.1
0.2 K
1 Digit
NTC 30
- 40 ... 0
0.1
1.0 K
1 Digit
0 ... 125
0.1
0.3 K
1 Digit
Measuring input
(μA):
Karl Fischer (Dead-Stop) connector (µA) for double platinum electrode.
Polarisation voltage variably adjustable from 40 … 220 mV.
Connector: 2 x 4 mm
– sockets.
Measurement
range I [
A]
Display resolution
Measurement accuracy* without
sensor probe
100
0.1
-5 /+ 3
A
1 Digit
50
0.1
+/- 3
A
1 Digit
10
0.1
+/- 1
A
1 Digit
5
0.1
+/- 0.2
A
1 Digit
* The measurement uncertainty of the sensor probe has to be taken into account as well.