SonTek – a Xylem brand
FlowTracker2 User’s Manual (January 2019)
64
4.5. Automated Beam Check
The
Automated Beam Check
function allows the user to perform a Beam Check in the
region of the measurement section before data collection starts. The automated beam
check perform a number of quality control checks on the data collected to determine if
the flow conditions are suitable for discharge measurements.
The
Automated BeamCheck
procedure and data analysis is discussed in detail in
of the manual. It is recommended that the user study this
section to get a full understanding of the BeamCheck procedure.
The
Automated Beam Check
screen shows the
steps involved in performing an automated beam
check.
a)
The FlowTracker2 probe should be placed in
the region of the measurement section in
moving water such that the probe is
submerged and well away from any
underwater obstacles.
b)
The FlowTracker2 collects data for about 20
seconds.
Figure 4:9 - Automated Beam
Check
c) The graphical display of the Quality Control parameters supplied during the
Automated BeamCheck is time series of each parameter.
d) The display options for individual or all beam check data are listed in
The shortcut key on the keypad determines which beam SNR will be displayed,
Table 4-6 - Automated Beam Check Display
Beam
Line Color
Keypad Key
Beam1
Red
1
Beam2
Blue
2
Beam3
Green
3
All Beams
All
4
e)
To select a Quality Control Criteria,
i).
Use up or down scroll arrows keys to view quality control criteria.
f)
To start the automated beam check,
i).
Press the Left Soft Key.
g)
To
CANCEL
automated beam check,
i).
Press the Right Soft Key,
ii).
The software will navigate to Utilities window.
h)
To accept Automated Beam Check,
Содержание SonTek FlowTracker2
Страница 107: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 106 ...
Страница 117: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 116 ...
Страница 196: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 195 ii Stationary Measurement ...
Страница 198: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 197 Measurement File Framework 12 1 1 ...
Страница 216: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 215 Variable Mode Description measurement ...
Страница 247: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 246 Configuration Changes ...
Страница 280: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 279 Figure 14 36 Edit by Station Edit a Station ...
Страница 283: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 282 ...
Страница 299: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 298 Appendix A Software Flow Diagram ...
Страница 300: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 299 ...