SonTek – a Xylem brand
FlowTracker2 User’s Manual (January 2019)
60
The
Raw Data
screen includes the following
variables,
a) Velocity,
b) SNR,
c) Temperature,
d) Tilt,
e) Battery,
f) Pressure.
g)
To select a variable,
i).
Use up or down scroll arrows keys to view a
variable.
h)
To navigate to Utilities menu,
i).
Press the Left Soft Key.
Figure 4:7 - Raw Data
Raw Data displayed in this function are NOT stored but enable the user to verify if the
flow conditions are ideal for measurements
.
Velocity Raw Data
4.4.1
The
Velocity
variable displays the individual velocity components of the one second
measured raw velocity data.
a) Velocity data can be expected to show notable variations (most of which are real),
and should be indicative of the general conditions in the water,
b) The raw velocity data are updated once per second,
c) The FlowTracker2 Probe Coordinate System is outlined in
Figure 4:8 - FlowTracker2 Coordinate System
d) The velocity components of FlowTraker2 and display options for individual or all
velocity components are listed in
. The shortcut key on the keypad
determines which velocity component will be displayed.
Table 4-3 - Raw Velocity Display
Sampling
Volume
Y
X
Probe
Coordinate System
Graduated Tag Line
Primary Flow
Direction
Mounting
Pin
Pressure
Sensor
Содержание SonTek FlowTracker2
Страница 107: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 106 ...
Страница 117: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 116 ...
Страница 196: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 195 ii Stationary Measurement ...
Страница 198: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 197 Measurement File Framework 12 1 1 ...
Страница 216: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 215 Variable Mode Description measurement ...
Страница 247: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 246 Configuration Changes ...
Страница 280: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 279 Figure 14 36 Edit by Station Edit a Station ...
Страница 283: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 282 ...
Страница 299: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 298 Appendix A Software Flow Diagram ...
Страница 300: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 299 ...