SonTek – a Xylem brand
FlowTracker2 User’s Manual (January 2019)
44
yellow.
g)
To navigate to Template Menu,
i).
Press the Left Soft Key.
Properties
associated with the parameters in data collection settings are defined in
Table 3-8 - Data Collection Settings - Properties
Parameter
Min
Max
Default
Units
Required
Averaging Time
10
1000
40
seconds
Yes
Salinity
0
42
0
PSU
Yes
Temperature
-5
50
Internal Measured
°C or °F
Yes
Sound Speed
1400
1580
Internal Calculated
m/s or ft/s
Yes
Mounting Correction
-5
5
0
%
Yes
Averaging Time
3.4.2.1
The
Averaging Time
parameter specifies the time duration (in seconds) of data
collection during each point velocity measurement.
a) Averaging time is specified in 1-second intervals from 10 to 1000 seconds.
b) The default Averaging Time applied in FlowTracker2 is 40s.
c)
To change the Averaging Time,
i).
Use the backspace key to clear current entry,
ii).
Type in the new value using the keypad.
The averaging time required for single point velocity measurement is usually from 40 to
60 seconds. This however is site dependent and longer averaging times are required
when measuring in turbulent flow conditions.
Salinity
3.4.2.2
The
Salinity
parameter specifies the salinity value in
Practical Salinity Scale
(dimensionless) used to compute sound speed.
a) Salinity is specified in Practical Salinity Scale (PSS-78). Fresh water has a
salinity of 0; seawater salinity generally ranges from 31 to 39 (PSS). The PSS-78
range is 2 to 42 for a temperature range of -2°C to +35°C and hydrostatic
pressure 0 – 10000dBar.
b) The default Salinity applied in FlowTracker2 is 0 (PSS).
Where a measurement location exceeds the PSS-78 range, Salinity cannot be used for
calculation of Sound Speed.
Sound Velocimeters
must be used to measure the actual
sound speed of the water directly at the measurement location.
c) As a rule of thumb, a 12 (PSS) error in the value of salinity will result in a 1% error
in sound speed, which results in a 2% error in velocity data.
Содержание SonTek FlowTracker2
Страница 107: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 106 ...
Страница 117: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 116 ...
Страница 196: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 195 ii Stationary Measurement ...
Страница 198: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 197 Measurement File Framework 12 1 1 ...
Страница 216: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 215 Variable Mode Description measurement ...
Страница 247: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 246 Configuration Changes ...
Страница 280: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 279 Figure 14 36 Edit by Station Edit a Station ...
Страница 283: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 282 ...
Страница 299: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 298 Appendix A Software Flow Diagram ...
Страница 300: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 299 ...