13.8. FlowTracker2 Data
Basic Sampling Strategy
13.8.1
The FlowTracker2 collects a burst of velocity data at each measurement location. The
sampling strategy for the FlowTracker2 instruments is summarized in
The
Original FlowTracker sampling strategy is included for comparison.
Table 13-1- Sampling Strategy
Category
FlowTracker2
Original FlowTracker
Sampling
Rate
Pings per
Sample
Ping Rate
Sampling
Rate
Pings per
Sample
Ping Rate
Field
2 Hz
20
40 Hz
1 Hz
10
10 Hz
a) An individual measurement of the 2D or 3D velocity is referred to as a ping.
b) Sample refers to the mean of 20 pings to produce a measurement of the 2D or
3D water velocity.
c) Two samples are recorded per second during the user specified averaging time.
d) The averaging time at each location is user-specified (10 to 1000 seconds).
e) The system collects a time-series of velocity at each measurement location over
the user specified averaging interval.
Velocity Data
13.8.2
The FlowTracker2 provides several important performance advantages.
a) It can measure 2D or 3D water velocities from 0.001 to 4.0 m/s (0.003 to 13 ft/s).
b) Velocity data are accurate to 1% of the measured velocity in a one-second
sample.
c) Velocity data can be used immediately without any post processing corrections.
d) True 2D or 3D velocity data are output in Cartesian coordinates (XYZ) relative to
probe orientation.
e) The FlowTracker2 calibration will not change unless the probe is physically
damaged. No periodic calibration is required.
f) Diagnostic software is included to evaluate system performance periodically.
Accuracy of Velocity Data
13.8.3
Optimizing of FlowTracker2
13.8.3.1
The FlowTracker2 has been optimized to provide the best possible velocity data.
a) Velocity data are accurate to 1% of measured velocity for each one-second
sample.
b) Variations in velocity data are dominated by true variations in water velocity.
c) Instrument noise does not normally have an impact on velocity data.
Содержание SonTek FlowTracker2
Страница 107: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 106 ...
Страница 117: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 116 ...
Страница 196: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 195 ii Stationary Measurement ...
Страница 198: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 197 Measurement File Framework 12 1 1 ...
Страница 216: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 215 Variable Mode Description measurement ...
Страница 247: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 246 Configuration Changes ...
Страница 280: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 279 Figure 14 36 Edit by Station Edit a Station ...
Страница 283: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 282 ...
Страница 299: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 298 Appendix A Software Flow Diagram ...
Страница 300: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 299 ...