![Xylem SonTek FlowTracker2 Скачать руководство пользователя страница 228](http://html1.mh-extra.com/html/xylem/sontek-flowtracker2/sontek-flowtracker2_user-manual_3888944228.webp)
SonTek – a Xylem brand
FlowTracker2 User’s Manual (January 2019)
227
Figure 13:2 - Signal to Noise Ratio Profile
13.4. Pulse-Coherent Processing
This section does not attempt to provide a detailed description of pulse-coherent
processing. It presents a simple overview with a focus on how this affects
FlowTracker2 operation. SonTek can provide additional information on request.
The description of FlowTracker2 operation given in
Bistatic Doppler Current Meters
an oversimplification.
•
Bistatic Doppler Current Meters
describes incoherent Doppler processing in
which the transducer sends a single pulse of sound and measures the frequency
change of the return signal.
•
The FlowTracker2 uses a technique called pulse-coherent processing.
Pulse-coherent processing provides the best possible performance of any Doppler
processing technique. In the simplest terms, pulse-coherent processing works as
follows.
a) The FlowTracker2 sends two pulses of sound separated by a time lag (
τ
).
b) Each receiver measures the phase (
φ
) of the return signal from each pulse.
c) The change in phase (
φ
2 -
φ
1) divided by the time lag (
τ
) is proportional to
velocity.
d) Because of the nature of the phase measurement (which can only be determined
from -180
°
to +180
°
), the system has an inherent maximum velocity limitation.
e) The FlowTracker2 velocity algorithms have been optimized to give the best
possible performance over a wide velocity range of
±
4.0 m/s (
±
13 ft/s).
f) FlowTracker2 processing provides unmatched results for low flows (<1 cm/s;
<0.03 ft/s).
FlowTracker2 processing has been designed to give the best possible performance in
all environments. However, there is a situation where system performance may be
affected by operating conditions.
Time after transmit pulse
R
ec
ei
v
ed
s
ig
n
al
s
tr
en
g
th
Center of
Receiver Axis
Ambient
Noise Level
Содержание SonTek FlowTracker2
Страница 107: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 106 ...
Страница 117: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 116 ...
Страница 196: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 195 ii Stationary Measurement ...
Страница 198: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 197 Measurement File Framework 12 1 1 ...
Страница 216: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 215 Variable Mode Description measurement ...
Страница 247: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 246 Configuration Changes ...
Страница 280: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 279 Figure 14 36 Edit by Station Edit a Station ...
Страница 283: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 282 ...
Страница 299: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 298 Appendix A Software Flow Diagram ...
Страница 300: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 299 ...