![Xylem SonTek FlowTracker2 Скачать руководство пользователя страница 226](http://html1.mh-extra.com/html/xylem/sontek-flowtracker2/sontek-flowtracker2_user-manual_3888944226.webp)
SonTek – a Xylem brand
FlowTracker2 User’s Manual (January 2019)
225
b) The velocity, V, represents the relative speed between source and receiver
(motion that changes the distance between the two).
i). If the distance between the two objects is decreasing, frequency increases.
ii). If the distance is increasing, frequency decreases.
iii). Motion perpendicular to the line connecting source and receiver does not
introduce a Doppler shift.
13.3. Bistatic Doppler Current Meters
Bistatic Doppler Operation
13.3.1
The FlowTracker2 is a Bistatic Doppler current meter with the 2D side looking probe
design and sampling volume shown in
a) Bistatic means separate acoustic transducers are used for transmitter and
receiver.
b) The transmitter generates sound concentrated in a narrow beam.
c) The receivers are sensitive to sound coming from a narrow beam.
d) The receivers are mounted such that the beams intersect at a volume of water
located a fixed distance (10 cm; 4 in) from the tip of the probe.
e) The beam intersection determines the location of the sampling volume (the
volume of water in which measurements are made).
The FlowTracker2 Measurement Principle
13.3.2
a) The transmitter generates a short pulse of sound at a known frequency.
b) The sound travels through the water along the transmitter beam axis.
c) As the pulse passes through the sampling volume, sound is reflected in all
directions by particulate matter (sediment, small organisms, bubbles).
d) Some portion of the reflected energy travels back along the receiver beam axes.
e) The reflected signal is sampled by the acoustic receivers.
f) The FlowTracker2 measures the change in frequency (Doppler shift) for each
receiver.
g) The Doppler shift is proportional to the velocity of the particles along the bistatic
axis of the receiver and transmitter. The bistatic axis is located halfway between
transmit and receive axes.
h) Knowing the relative orientation of the bistatic axes allows the FlowTracker2 to
calculate 2D or 3D water velocity.
Содержание SonTek FlowTracker2
Страница 107: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 106 ...
Страница 117: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 116 ...
Страница 196: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 195 ii Stationary Measurement ...
Страница 198: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 197 Measurement File Framework 12 1 1 ...
Страница 216: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 215 Variable Mode Description measurement ...
Страница 247: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 246 Configuration Changes ...
Страница 280: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 279 Figure 14 36 Edit by Station Edit a Station ...
Страница 283: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 282 ...
Страница 299: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 298 Appendix A Software Flow Diagram ...
Страница 300: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 299 ...