SonTek – a Xylem brand
FlowTracker2 User’s Manual (January 2019)
213
Variable
Mode
Description
%Dep
D, G
Measurement depth location, as fraction of
the effective depth
MeasD
D, G
Point measurement depth, measured form the
water surface
Npts
D, G
Number of samples in measurement; one
sample collected per second
Spike
D, G
Number of spikes removed before calculating
mean values
VelX
D, G
Mean X velocity component of point velocity
measurement
VelY
D, G
Mean Y velocity component of point velocity
measurement
VelZ
D, G
Mean Z velocity component of point velocity
measurement
Angle
D, G
Velocity angle calculated as atan (Vy/Vx)
VxErr
D, G
Standard error of X velocity
VyErr
D, G
Standard error of Y velocity
VzErr
D, G
Standard error of Z velocity
SNR1
D, G
Signal-to-Noise Ratio beam 1
SNR2
D, G
Signal-to-Noise Ratio beam 2
SNR3
D, G
Signal-to-Noise Ratio beam 3
Bnd
D, G
Boundary QC measurement
Temp
D, G
Water temperature
CorrFact
D, G
Correction factor used to scale mean station
velocity
Tilt
D, G
Mean wading rod angle during point velocity
measurement
MeanV
D
Mean panel velocity based on panel discharge
and area.
G
Mean V is the mean of the X velocity of all point
velocity measurements.
Area
D
Station area
Flow
D
Station discharge
%Q
D
Percent of total discharge
The
Summary
file consists of the velocity method, velocity, SNR and standard error data
of each point velocity measurement. The variables exported in the summary file are
given in
Table 12-6 - Summary File Variables
Variable
Mode
Description
Desktop_Software_Build_Version
D, G
Version of desktop software
#
D, G
Burst or measurements number
Содержание SonTek FlowTracker2
Страница 107: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 106 ...
Страница 117: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 116 ...
Страница 196: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 195 ii Stationary Measurement ...
Страница 198: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 197 Measurement File Framework 12 1 1 ...
Страница 216: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 215 Variable Mode Description measurement ...
Страница 247: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 246 Configuration Changes ...
Страница 280: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 279 Figure 14 36 Edit by Station Edit a Station ...
Страница 283: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 282 ...
Страница 299: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 298 Appendix A Software Flow Diagram ...
Страница 300: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 299 ...