82
Fig. 69
Under «
KF Trac
e» the titrator to be used for the titration is selected. In the «
Oven
» section, the oven to be used
is selected. Depending on the connected system, the TO 7280 or the TW 7650 can be selected.
3.3.2
Working with the TO 7280
After selecting the oven, the additional oven options open (Fig. 70).
Fig. 70
You have the following options:
«
Gas source
»:
Selection between pump (ambient air) or gas (nitrogen).
«
Oven temperature
»:
Setting the working temperature of the oven.
«
Automatic fan control
»:
When the automatic fan is switched on, an additional fan switches on to
accelerate cooling processes.
«
Oven temperature Delta
»: Delta criterion above which the temperature reached is accepted.
«
Keep head down
After completion of a titration, the needle head remains down so that the
after method
»:
titrated sample vial can be used to measure a new base drift.
For general setup of a worklist see the TitriSoft .