84
Technical Specifications
1.3
1.3.1
Titrator TitroLine
®
5000
Translation of the legally binding German version
(Release: 18. June 2020)
EMC compatibility according to the Council Directive: 2014/30/EU;
applied harmonized standards: EN 61326-1: 2013
Low-voltage directive according to the Council Directive 2014/35/EU;
Testing basis EN 61 010-1: 2010 for laboratory equipment
RoHS Council Directive 2011/65/EU
FCC Part 15B and ICES 003
Country of origin
: Germany, Made in Germany
The following solvents/titration reagents are allowed to be used:
All common titration solutions.
As reagent water and all non-aggressive non-organic and organic fluids are allowed.
If using
combustible fluids fire please adhere to the Guidelines for Explosion Protection and Prevention of the
chemical industry.
For fluids with higher viscosity (
5 mm
2
/s), lower boiling point or affinity to outgas, the filling and dosage
speed can be adjusted.
Fluids with viscosity over 20mm
2
/s cannot be dosed.
To ensure maximum accuracy of the readings we recommend to allow some reasonable time for the
TitroLine
®
5000
to “warm up”.
Measuring input 1
(analog):
pH/mV-input with 12 bit transducer for high-precision readings.
Electrode socket according to DIN 19 262 or additional with BNC socket insert (Z 860).
Reference electrode 1 x 4 mm socket.
Measurement range Display resolution
Measurement accuracy*
without sensor probe
Input
resistance
pH
pH
- 3.0 ... 17.00
0.01
0.05
1 Digit
> 5
10
12
mV
U
mV
- 1900 ... 1900
1
1.0
1 Digit
> 5
10
12
Measuring input
(Pt 1000):
Temperature sensor - connector for a resistance thermometer Pt 1000 and NTC 30 kOhm.
Connection: 2 x 4 mm sockets.
Measurement
range T [°C]
Display resolution
Measurement accuracy* without
sensor probe
Pt 1000
- 30 ... 115
0.1
0.5 K
1 Digit
NTC 30
- 30 ... 115
0.1
0.5 K
1 Digit
Display:
3.5 inches -1/4 VGA TFT display with 320x240 pixels.
Calibration
:
Automatically with up to three buffer solutions, sequence during calibration optional, freely
definable buffers can be input.
Default buffer solutions according to DIN 19 266 and NBS, or technical buffers:
pH = 1.00; pH = 4.00; pH = 4.01; pH = 6.87; pH= 7.00; pH = 9.18; pH = 10.00
Input
:
Measurement input 1: pH/mV-input with electrode socket according DIN 19 262/or BNC
Measurement input Pt 1000: Temperature sensor probe for resistance thermometer
Pt 1000/NTC 30 (Connection sockets: 2 x 4 mm)
Power supply
:
power supply 100-240 V; 50/60 Hz, power input: 30 VA
Use the power supply TZ 1853 only!
* The measurement uncertainty of the sensor probe has to be taken into account as well.