pHix® Compact V3. NG (next generation)
EN 4.10 Version 2202
Mounting /Mechanical mounting
12
1.
2.
3.
4.
Check that the connectors are absolutely clean and dry.
Lubricate the O-ring in acid-free grease or Vaseline.
Check that the O-ring is placed correctly in the recess of the new electrode.
Check that the hole for the electrode in the pHix® Compact is completely dry – if
necessary use a dry cloth.
Screw in the new electrode and tighten by hand.
Note!
Do
not
use tools!
5. Finally, pHix® Compact must be
Mechanical mounting
pHix® Compact is made from PPS Fortron and/or stainless steel (EN 1.4404 / AISL316L)
and because of this, it has a tremendous resistance against chemicals. pHix® Compact
can, when mounted on a pipe socket, be used for measuring in solvents and strong
alcalic or acid liquids
Open and closed systems
pHix® Compact is designed for measuring in both open and closed systems.
Open systems are, for example, gutters, wells, and reservoirs. Closed systems are, for
example, pipe systems or tanks/vessels.
Содержание MJK pHix Compact V3 NG
Страница 1: ...Manual EN 4 10 Version 2202 pHix Compact V3 NG NEXT GENERATION...
Страница 2: ...pHix Compact V3 NG next generation EN 4 10 Version 2202 2 Your notes...
Страница 15: ...pHix Compact V3 NG next generation EN 4 10 Version 2202 Mounting Mechanical mounting 15...
Страница 85: ...pHix Compact V3 NG next generation EN 4 10 Version 2202 pHix Configurator software Online data 85 Status...
Страница 90: ......
Страница 91: ......