IQ SensorNet MIQ/IC2 Module

QUICK START GUIDE 

Quick Start Guide

YSI Municipal Water

 

 

XA00166

Overview 

The MIQ/IC2 current input module provides two 0/4 - 20 mA current 
inputs for IQ SensorNet and thus enables to connect external sensors 
via their current output. Measured values of the external sensors can 
be displayed, recorded and processed like the measured values from 
IQ SensorNet.

Examples:

 

•  Connecting flow meters to IQ SensorNet
•  Connecting level meters to IQ SensorNet
•  Connecting analyzers to IQ SensorNet 
•  Connecting pressure sensors to IQ SensorNet 

Scope of Delivery:

•  MIQ/IC2 module
•  4 x cable glands (clamping range 4.5-10 mm) with seals and blind 

plugs

•  4 x ISO blind nuts M4 with suitable cheese-head screws and plain 

washers

•  2 x countersunk screws M3x6 to close the module lid (+ 2 

replacement screws)

•  1 x contact base with fixing screws

Materials Required

To set up the MIQ/IC2 module you will need the following tools:
•  Cable stripping knife
•  Wire stripper
•  Phillips screw driver
•  Small screw driver
•  Cable

TABLE OF CONTENTS:

1. Setup
2. Wiring 
3. Configuration

Figure 1: 

MIQ IC2 Module

Содержание IQ SensorNet

Страница 1: ...ting level meters to IQ SensorNet Connecting analyzers to IQ SensorNet Connecting pressure sensors to IQ SensorNet Scope of Delivery MIQ IC2 module 4 x cable glands clamping range 4 5 10 mm with seals and blind plugs 4 x ISO blind nuts M4 with suitable cheese head screws and plain washers 2 x countersunk screws M3x6 to close the module lid 2 replacement screws 1 x contact base with fixing screws M...

Страница 2: ... below 2 Connect the cable to the terminal strip 1 Open the MIQ IC2 module 2 Open the cable gland fitting under the required input Keep the blind plug fitting for later modifications if necessary 3 Loosen the coupling ring 4 Feed the cable through the cable gland in the module housing 5 Open the terminal strip with a small screw driver 6 Connect the cable to the terminal strip with a small screw d...

Страница 3: ...onnection without powering the device The current outputs of external measuring systems can be directly connected to the current inputs of the MIQ IC2 module If the connection data of the external meter is suitable Apply power to device Keep in mind the following information when setting up the MIQ IC2 Module and third party devices IQSN Power Consumption The power consumption is 0 2 watts plus 2 ...

Страница 4: ...Scroll down to the Setting of sensors and diff sensors option and press OK Select the Measuring range column and press OK 7 Press Continue Customize the following settings to the appropriate values 8 5 6 ...

Страница 5: ...5 Measuring range Decimal places 11 Display value 0 4mA Display value 20 mA 12 9 10 ...

Страница 6: ...6 Display unit Measured parameter 15 Error detection Error threshold 16 13 14 ...

Страница 7: ...ings OH 45387 1 937 767 7241 info ysi com YSI com IQSN 2021 Xylem Inc XA00166 0621 QUICK START GUIDE XA00166 MIQ IC2 REC 2 If only one device is connected to MIQ IC2 module Inactive If two devices are connected to MIQ IC2 module Active Press Save and quit 18 17 ...

Отзывы: