Checking the zero point of the sensor 

3.4

There are two methods to check the zero point of the sensor: 

  Measurement in a nitrogen atmosphere (recommended method). 

  Measurement in a sodium sulfite solution according to DIN EN 25814/ ISO 5814. 

 

Test criterion

 

The sensor is OK if the measuring instrument displays < 0.5 % D.O. 
saturation after 15 minutes. 
 

 
 

What to do if ... 

Error symptom

 

Cause

 

Remedy

 

No temperature value or 
D.O. value

 

 No connection between meter and 

D.O. sensor 

 Establish connection between 

meter and D.O. sensor 

 Cable defective 

 Return the D.O. sensor 

Measured value too high 
or too low or error 
message: 

Error

 

 Coating on sensor cap 

 Clean the outside of the sensor  

(

 chapter 3.3.1) 

 Membrane damaged 

 Exchange the sensor cap 

 Service life of the sensor cap over 

 Check the sensor 

 Replace the sensor cap as 

necessary (

 Chapter 3.2) 

 Dirt inside the sensor cap and 

sensor head. 

 Protective hood not properly 

tightened. 

 Sensor cap untight or defective. 

 Dismantle the sensor cap 

 Clean the insides of the sensor cap 

and sensor head (

 Chapter 

3.3.2) 

 Mount the sensor cap correctly and 

tighten the protective hood as far 
as it will go (

 Chapter 3.2) 

 Replace a defective sensor cap as 

 

Incorrect temperature 
display

 

 The temperature sensor is not 

immersed deep enough in the 
measuring solution 

 Observe the minimum immersion 

depth 

 Temperature sensor defective 

 Return the D.O. sensor 

Error message: 

no cap

 

 No sensor cap put on 

 Clean the sensor head and sensor 

cap 

 Sensor cap not recognized 

 Return the D.O. sensor 

 Sensor cap defective 

 Replace the sensor cap 

 

 

Содержание FDO 925/STM

Страница 1: ...2020 03 11 Operating Manual FDO 925 STM Optical Dissolved Oxygen Sensor ...

Страница 2: ......

Страница 3: ...ns for use are intended to show you the proper and safe handling of the product For the highest possible safety you must adhere to the listed safety and warning notes in these manual The pictograph has the following meaning Warning of a general danger No compliance results can result in injury or material damage Important information for the use of the device Refers to another part of the operatin...

Страница 4: ...nics 3 Connection cable FDO 925 STM 4 FDO converter 5 Schutzkorb 6 Sensor membrane 7 Sensor cap SC FDO 925 with memory chip 8 Gold plated contacts for the memory chip of the sensor cap 9 Locking device 10 Measurement window 11 O ring 12 Thermistor enclosure with temperature sensor 1 2 3 6 7 8 10 11 12 5 4 9 ...

Страница 5: ...matically retained if the sensor or meter are exchanged The digital transmission technique guarantees the failure free communication with the meter even with long connection cables Recommended fields of application 1 2 On site measurements in rivers lakes and wastewater Applications in water laboratories BOD measurements 2 Measuring Operation General information on the handling of the sensor cap 2...

Страница 6: ... required When does a check or user calibration make sense A function check or user calibration can be useful in the following special cases If the measured values appear to be implausible and it is assumed that the service life of the sensor cap is over Routinely within the framework of the company quality assurance Check or calibration medium The check and user calibration take place in water va...

Страница 7: ...interior of the sensor cap will by and by affect the measurement characteristics and shorten the service life of the sensor cap Therefore the interior of the sensor cap should not be exposed to direct sunlight Avoid any exposure to light that exceeds the extent required for necessary maintenance and cleaning activities Store dismantled sensor caps in a light protected environment only Exchanging t...

Страница 8: ...e sensor with tapwater to remove loosely adhering dirt Rough dirt on the sensor shaft can be brushed off with a soft brush Attention Do not use the brush in the area of the sensor membrane Risk of damage The sensor cap including the sensor membrane should be wiped with a soft and moist microfiber cloth In the case of persisting dirt you can add some household washing up liquid to the tapwater Atte...

Страница 9: ...nsor chapter 3 3 1 Membrane damaged Exchange the sensor cap Service life of the sensor cap over Check the sensor Replace the sensor cap as necessary Chapter 3 2 Dirt inside the sensor cap and sensor head Protective hood not properly tightened Sensor cap untight or defective Dismantle the sensor cap Clean the insides of the sensor cap and sensor head Chapter 3 3 2 Mount the sensor cap correctly and...

Страница 10: ...hPa 0 5 of measured value Temperature 0 50 0 C 0 2 C Membrane shape chamfered Shaft material POM stainless steel Shaft length 140 mm 1 mm Diameter 15 3 mm 0 2 mm Cable length 10 m Protection IP 68 Converter Output signal 2 x 4 20 mA Measuring range 0 00 20 00 mg l DAC0 4 20 mA 0 0 50 0 C DAC1 4 20 mA Error detection no data 0 mA no cap 22 mA Length 165 mm 10 mm Diameter 16 mm 1 mm Dimension in mm ...

Страница 11: ...ges Developing new technologies that will improve the way water is used conserved and re used in the future is central to our work Our products and services move treat analyze monitor and return water to the environment in public utility industrial residential and commercial building services settings Xylem also provides a leading portfolio of smart metering network technologies and advanced analy...

Отзывы: