Installation
DIQ/S 281
ba76226e02
03/2023
28
Connecting the
cables
The SNCIQ and SNCIQ/UG cables are connected to the terminal strip in the
same way as the SACIQ sensor connection cable (see section 3.5.4):
1
Open the module (DIQ, MIQ).
2
Select any SENSORNET connection. At the same time, look out for the
SENSORNET designation.
figure 3-12 Connecting cables to the MIQ module
1
Cable gland
2
Seal
3
Coupling ring
3
Screw the cable gland (pos. 1 in figure 3-12) with the sealing ring (pos.
2 in figure 3-12) into the module housing.
4
Loosen the cap nut (pos. 3 in figure 3-12).
5
Feed the cable through the cable gland into the module housing.
6
Connect the cable ends to the terminal strip. At the same time, look out
for the designations of the terminals (red / shield / green).
SENSORNET 2
SENSORNET 1
SACIQ
SNCIQ
SNCIQ/UG
1
3
2
X6
X5
X4
SENSORNET 2
X3
X2
X1
SENSORNET 1
RED
RED
SHIELD
SHIELD
GREEN
GREEN
ON
OFF
SN
TERMINA
TO
R