Special features menu 

Setting up your OPTi

Your OPTi has been delivered with “Brix” in scale position A (

S.A.01

).  For many applications this is enough to start working straight out of the box. 

When first switching on, the instrument will briefly display the scale loaded to  
each channel.  As there is only one scale active, the initial display will be:

Opti   

>>

S.A.01

To overwrite and/or load other scales in A, B and C, the instrument’s special features menu must be accessed. Once loaded, the scale list on start-up 
will expand to show the active scales:

Opti   

>>

S.A.01   

>>

s.b.34   

>>

s.c.22

Please note that some distributors may pre-configure your OPTi so that the scales you require are correctly installed.  If so, please skip this initial 
process.

Accessing the special features menu

To access the special features menu:
1.  Switch on the OPTi
2.  Press and hold READ until  -

OFF

- is displayed

3.  Whilst 

-Off-

 is still displayed, quickly release  

      READ and press and release ZERO

Navigate through the menu and configure the instrument using the ZERO 
and READ keys.
•  ZERO = Step through the menu
•  READ = Enter a function 
•  ZERO = Step though the function options  
    (if applicable)
•  READ = Set the option and exit the feature

Display Feature

Access

Actions

-Ag.t-

AG Test Mode

Open

Press READ to switch on AG Test Mode
The instrument will enter the AG Test Mode
Switch the instrument OFF to exit the AG Test Mode

-s.a.- 

Scale A

Open or PIN 

Press ZERO to step through the list and find a scale
Press READ to set and return to the menu

-s..b.- 

Scale B

-sc..-

Scale C

-t.Sc-

Temperature display 

Open or PIN 

Press ZERO to step through °F or °C
Press READ to set and return to the menu

-dLy-

Delay before Read/
Zero

Open or PIN 

Press ZERO to step through 0 to 60 in 5-second intervals
Press READ to set and return to the menu

-Pin-

Set /Remove PIN 
protection

Openor PIN 

To set your PIN:

1. Press ZERO to increment the flashing digit
2. Press READ to select the number and move to the next digit
3. Repeat to confirm
4. If correct then 

-set-

 is displayed

5. The instrument will leave the menu

To remove PIN protection:

1. Enter the previously set PIN (see steps 1 and 2 above)
2. If the entered PIN is correct 

PASS

 is displayed and  

    the PIN is removed
3. The instrument will return to the menu

-Uer-

Version

Open

View software version and serial number

-End-

Save/Exit

Open

Save settings and leave the menu

This instrument has been calibrated and checked for precision at 20°C.  
Analytical Grade Water

1

 and AG Calibration Fluids

2

 were employed to check the accuracy of the instrument for the primary scales.  All other scales are 

assumed as correct as their results are  calculated emperically from these values.

Sample Type

Certified Value (Brix/RI)

Acceptance Limits

Analytical Grade Water

0.00 / 1.33299

±0.2 Brix / ±0.0003 RI

AG30 Fluid

30.00 / 1.38115

AG50 Fluid

50.00 / 1.42009

Certificate of verification

All materials used to verify the performance of this OPTi refractometer were callibrated by Bellingham & Stanley Ltd., UKAS accredited calibration laboratory no. 0834, accredited to ISO/
IEC 17025:2017. 
1.  Refractive index values for Water are obtained from "Revised Formulation for the Refractive Index of Water and Steam as a Function of Wavelength, Temperature and Density", adopted 
by the International Association for the Properties of Water and Steam (IAPWS) and available as part of NIST Standard Reference Database 10. Refractive indices calculated from the 
formulation are absolute refractive indices; conversion to refractive index against air requires division by the respective absolute refractive index of air (NIST Engineering Metrology 
Toolbox).
2.  AG calibration fluids are prepared by mixing water with a soluble chemical of higher refractive index. The Brix value of the fluid is determined using a high accuracy digital 
refractometer at 20.0°C, which has been calibrated with a NIST traceable sucrose calibration standard of Brix value equal to the target value for the AG fluid according to a Laboratory 
Procedure  
QL-103. The AG fluid composition is finely adjusted to give a measured value to within ± 0.01 Brix (± 0.00002 RI) at 20.0°C. 