Page 44
June 2022 – TD 320 OPERATING MANUAL – TURBIDITY SENSOR 4296
3.5.2 Terminal Protocol
The properties in this menu are highly dependent on the mode setting. If sensor is set to
AiCaP
these properties have no influence on the sensor. In
AADI Real-Time
mode only
Enable Polled
mode
will be active. In
Smart Sensor Terminal
and
Smart Sensor Terminal FW2
mode all three
properties are used.
If the
Enable Polled Mode
is enabled,
then the sensor outputs data when the user/system polls
for data with a
Do Sample
command.
Enable Text
and
Enable Decimalformat
control the output string in Smart Sensor Terminal.
With
Enable Text
enabled the sensor will output a string with parameter name together with each
reading, refer
chapter 5.8.4
for an example where this command is toggled.
Enable Decimalformat
toggle between decimal format like 0.10 and Engineering format like
1.000E-01.
Figure 3-21: Terminal Protocol settings in System Configuration