![Xylem 627379-01 Скачать руководство пользователя страница 31](http://html1.mh-extra.com/html/xylem/627379-01/627379-01_user-manual_892016031.webp)
31
Read/Write Multiple Registers
Address
Description
R/W
Type
Default
0x0100
ODO Cap Coefficient K1 [3:0]
RW
U32
N/A
0x0101
0x0102
ODO Cap Coefficient K2 [3:0]
RW
U32
N/A
0x0103
0x0104
ODO Cap Coefficient K3 [3:0]
RW
U32
N/A
0x0105
0x0106
ODO Cap Coefficient K4 [3:0]
RW
U32
N/A
0x0107
0x0108
ODO Cap Coefficient K5 [3:0]
RW
U32
N/A
0x0109
0x010A
ODO Cap Coefficient K6 [3:0]
RW
U32
N/A
0x010B
0x010C
ODO Cap Coefficient K7 [3:0]
RW
U32
N/A
0x010D
0x010E
ODO Cap Coefficient KC [3:0]
RW
U16
N/A
0x010F
ODO Cap Replacement Time (s) [3:0]
RW
U32
N/A
0x0110
0x0210
ODO Last Calibration Time (s) [3:0]
R
U32
N/A
0x0211
0x0212
ODO Calibration QC Score
R
U16
N/A
0x0120
ODO Cap Serial Number [0:8] + pad
R
ASCII
N/A
0x0121
0x0122
0x0123
0x0124
0x0220
ODO Zero Calibration Time (s) [3:0]
W
U32
N/A
0x0221
Modbus Configuration & Commands
0x0230
ODO % Calibration Time (s) [3:0]
W
U32
N/A
0x0231
0x0232
ODO % Calibration Barometer (mmHg) [3:0]
W
Float
N/A
0x0233