Operating Instructions 

Immersion heating circulators 

UXP 8011-HC090 
UXP 8011-HC130 
UXP 8011-HC250 

Содержание UXP 8011-HC090

Страница 1: ...SHANGHAI 200032 CHINA TEL 86 21 6443 1641 FAX 86 21 6403 9005 EMAIL BC XTEMP USA CN XTEMP ENVIRONMENTS SHANGHAI CO LTD ADD ROOMS 909 11 NO 2777 ZHOUHAI ROAD PUDONG NEW DISTRICT SHANGHAI 200137 CHINA T...

Страница 2: ...it 14 2 2 5 Accessory cooling coil Optional 15 2 2 6 Put into operation 15 Chapter 3 Controller Descriptions and Operation 17 3 1 Controller panel 17 3 1 1 Display for temperature and working paramete...

Страница 3: ...08 2 V211119 UXP8011 CD6400 CLASS I NFL RT 5 100 C Contents C 4 2 Regular check of the STB 27 4 3 Checking and replacing the fuse 28 4 4 Clear error warning and alarm 29 Chapter 5 Equipment Overview a...

Страница 4: ...trical energy The operator is largely protected through the application of the appropriate standard specifications Further hazard sources may arise from the type of material being treated e g mixing d...

Страница 5: ...he conditions that the equipment is operated and serviced exclusively in accordance with its intended purpose and as described in these Operating Instructions the equipment is not modified only origin...

Страница 6: ...egrade electric safety of the equipment Note the effect of the altitude on the boiling point of the heat transfer liquids 1 2 Safety warning When heating and conditioning with specimen or application...

Страница 7: ...clear of any obstacles Keep the equipment away from hot air emitting source direct sunshine strong magnetic field and electric sparks Neither contamination of high concentration dust or corrosive gas...

Страница 8: ...iquids Use special care when using with these flammable liquids including the proper setting of the adjustable overheating protection device The circulation of the heat transfer liquid is important fo...

Страница 9: ...kink and other mechanical stress develops during normal use which otherwise may result in thermal and or liquid hazards 2 Flexible hose fitting and coupling are important for the safe operation of the...

Страница 10: ...azard 2 Don t touch the frozen surface or any parts with cold residual liquid Warning Hazard of frostbite Caution Hazard of possible electric shock when removed and placed in horizontal or up side dow...

Страница 11: ...servicing or repairing Accessories lists UXP8011 HC090 Part Number Qty Descriptions Remarks DSGO0708 2 1 Operating instructions This document ERSG0202 1 Fuse 250V 15A D5L20 Breaking capacity 1500A BJG...

Страница 12: ...f the equipment with protective ground PE For possible wet location conditions where cycle of evaporating and condensing overflowing or spillage may occur and for protection against hazard of possible...

Страница 13: ...ment Use only hose fitting and coupling provided with the equipment or recommended by Xtemp The ratings of pressure and temperature for the hose fitting and the jacket container or heat exchanger of t...

Страница 14: ...stacles so that no close loop of the air circulating may develop If no external circulating is necessary use hose fitting and coupling to short circuit the liquid connections as for an external applic...

Страница 15: ...ne adaptor SL The connection to the pump outlet and external circulating kit shall be hydraulically tight Loose connection may result in hazard of liquid spray The pump inlet shall be clear of any obs...

Страница 16: ...coil Optional Fig 5 Cooling coil For immersion heating bath GP series and circulators completed with open bath tank the minimum operating temperature of the equipment is limited to 5 10 C above ambien...

Страница 17: ...th non flammable and flammable heat transfer liquids Alcohol and silicone oils are the most common flammable heat transfer liquids Use special care when using with these flammable liquids including th...

Страница 18: ...mit Ht The temperature setting beyond the scope will not be accepted Ht This parameter is an absolute value and used for reminding only When temperature in the equipment is higher than the SEt value H...

Страница 19: ...Auto Select between AoFF and Aon which means manually starting and automatically starting Atc Absolute temperature calibration ATC when the temperature displayed differs from the temperature measured...

Страница 20: ...r may be accompanied by the following Ht When temperature in the bath is higher than the set value Ht is displayed Lt When temperature in the bath is lower than the set value Lt is displayed Er01 Warn...

Страница 21: ...the digits or setting by pressing accept the setting by pressing or confirm and exit automatically in 4 seconds 5 Set the programming parameters Press and at the same time then release the buttons 1...

Страница 22: ...en the equipment is in running mode timer for ASd is valid while the equipment is in standby mode timer for AS is valid The time is set by AS and ASd 8 Move between settings or increase the value Pres...

Страница 23: ...ipment is in operation and start to heat When the set temperature and actual temperature are quite different heating LED will be steadily lit with full power heating and the temperature will go up wit...

Страница 24: ...liquid The first time commissioned for use or when exchange of the heat transfer liquid and immediately after startup of the circulating pump check if any liquid leakage exists in the external circul...

Страница 25: ...priate for calibration 3 5 Setting for restart mode after power failure Auto Press for more than 2 seconds until screen displays Ht Scroll the menu by until Auto is displayed then press to enter Use t...

Страница 26: ...The function of temperature programming is optional 2 Press and at the same time and then release The parameter of first segment t1 is displayed Press to display original time press for a new value co...

Страница 27: ...or disengaging the residual current circuit breaker RCD Pull the temperature of the heat transfer liquid as close as possible to the ambient before attempting to shut down the equipment except when o...

Страница 28: ...onnect the power immediately Power on only after water or foreign material is completely evaporated or removed 4 2 Regular check of the STB Regular check of the STB is required for safe operation Inte...

Страница 29: ...capacity as rated and do as follows 1 Make sure the mains plug has been removed 2 Loosen the two screws on the top and 4 screws at the back of the controller as shown in Fig 7 Remove the cove plate an...

Страница 30: ...00 checked or replaced and for servicing or maintenance call authorized service engineer When the equipment displays Stb try first to pull down the temperature of the heater or the heat transfer liqui...

Страница 31: ...and microprocessor based precise dynamic temperature control with temperature stability up to 0 05 C Independent immersion controller head DIY for precise temperature system with different bath capac...

Страница 32: ...olute Temperature Calibration Auto start up or standby after power failure Duty cycle Continuous Wetted materials Stainless steel deep drawn Gastro norm pan Heater and pump of SUS316L Enclosure materi...

Страница 33: ...DSGO0708 2 V211119 UXP8011 CD6400 CLASS I NFL RT 5 100 C...