ForeSite Intelligent Surface System User Guide
Maintenance and Cleaning
XSENSOR Technology Corporation
30 of 46
Rinse off residual cleaning solution using a clean soft cloth that has been wetted with cool tap water by
wiping until all evidence of residual cleaning solution is removed.
2
O
H
H
2
O
Repeat steps 3–13, for a total of two bleach-rinse-wipe cycles.
Allow the system to dry completely.
Re-cover system with a sheet.
Caution: Follow the specific EPA label contact times when using the disinfectant.
Caution: Solvents containing alcohol, bleaching, or abrasive material may be harmful. Contact XSENSOR
before using any disinfectant not listed above.
Caution: Under no circumstances should a phenolic-based cleaning solution be used.
Warning: If the sensor becomes wet or soiled it should be cleaned immediately as described here.
Warning: Do not use excessive liquid or harsh cleaners. Do not use any steam cleaning device on the
sensor. The sensor can be damaged by excessive moisture. The bleach solution must not exceed 10% bleach.
Clorox Dispatch® or an equivalent bleach is recommended.
When finished cleaning, visually inspect the sensor for missed stains or soiling. Repeat cleaning process for any
missed spots. Allow to dry before storage or reuse.