11
UM Temp74 PT100 Vio EN rev.1 19.05.2023
•
Reference probe
Through this menu we can select the Reference Probe. In the right column of the Multiview screen, for each
channel, the difference of the measurement from the reference channel will be displayed.
•
Offset
The Offset menu allows you to adjust the measurement of each individual channel independently or to align all
measurements to the reference channel.
Offset adjustment (single channel)
•
Select which individual channel you wish to adjust (Probe 1-2-3-4-) or select Align All to align all readings to the
reference channel.
•
Choosing a single channel will open the screen allowing adjustment of the channel value in real time.
At the top the real-time reading is shown, at the bottom the adjustment value.
•
Using the buttons
change the 'Adjustment' value until the desired 'Current Temperature' is reached and
press
to confirm.
Note
: An adjustment of a maximum of ± 5.00 °C from the current temperature can be made.
If you select a channel with no probe connected, the display will show the error message: "CONNECT
THE PROBE", accompanied by the icon flashing
. If the offset has never been changed or a reset
has been carried out, the message ''Offset 0.00 °C'' will be shown for each channel.
Offset adjustment (align all)
This function allows the measurements of all channels to be aligned to the reference channel.
•
Select our reference channel from the Setup menu
Temp Settings -> Reference Probe and press
to
confirm.
•
In the Offset Probes menu, select Align All.
•
The measurements of all channels will be aligned to the measurement of the reference channel.
•
Calibration data
Accessing the Calibration Data menu will display the calibration data for each channel.
In addition to the value of the adjustment, the date and time at which it was made are displayed.
If no offset has been entered, 0.00 °C appears. The same screen is displayed each time it is switched on.
•
Resolution
The Resolution menu allows you to select how many decimal digits to show in measurement.
By default, the instrument is set with the highest resolution (0.01 °C).
Using the buttons
select the desired resolution and press
to confirm.
•
Stability filter
In order for a value reading to be considered true, it is advisable to wait for measurement stability, indicated by
the icon . Using the buttons
select the desired degree of stability and press
to confirm.
•
Low: with variations less than or equal to 3 hundredths of a degree, the stability icon remains fixed and
indicates the stable value.
•
Nor: with variations less than or equal to 2 hundredths of a degree, the stability icon remains fixed and
indicates the stable value.
•
High: with variations less than or equal to 1 hundredth of a degree, the stability icon remains fixed and
indicates the stable value.
+0,02
-0 ,05
+0,04
37,05 °C
37,07 °C
37,00 °C
37,09 °C
Multiview