15
•
Setup menu structure
P5.0 DO SETTINGS
P5.1 Cal 0
P5.2 Salt Compensation
P5.6 Cal Data
P5.8 Reset Settings
P5.9 Temp Cal
P9.0 SETTINGS
P9.1 Temperature U.M.
P9.3 Backlight mode
P9.4 Brightness
P9.5 Sleep Mode
P9.8 Reset Settings
P9.9 Auto Power-Off
Temperature measurement ATC – MTC
•
ATC
: The direct measurement of the sample temperature for all parameters is carried out through the
NTC 30KΩ probe, integrated into the sensor.
•
MTC
: If no temperature probe is connected, the value must be changed manually: keep pressed
or
until the value starts to flash; then adjust it by continuing to use the directional keys; press
to confirm.
Note
: with the polarographic sensor supplied together with the instrument, the manual temperature
compensation (
MTC
)
IS NOT TO BE PERFORMED.
%O
2
Parameter
Connect the polarographic sensor to the RCA / CINCH Temp connectors in the top panel of the device.
It is not necessary connect an external temperature probe, because it is already integrated.
•
O
2
parameter Setup
•
In measure mode press
to access the SETUP menu.
•
Press the button
to access the
DO SETTINGS P5.0
menu.
•
Move with the keys
and
to select the program to access.
The table below shows the setup menu structure for the O
2
parameter, and for each program the options