XS Instruments COND 7 Plus Скачать руководство пользователя страница 17

17

 

 

7+Series Version 2.0

 

 

temperature value blinks when press  

   key, press 

    and   

   key to adjust the temperature value, and 

when press  

    key, conductivity value blinks. 

 

 

 

Only 1-point calibration for customer-defined calibration. The value set in “customer-defined” is at a fixed 
temperature. There is no regulations of temperature coefficient and reference temperature. The meter has to perform 
calibration and measurement at the same temperature to avoid large error. The meter cannot recognize customer-
defined calibration solution.

 

 

As for the standard calibration also in customer calibration the meter show on the LCD one icon indicating the suited 
range.

 

Display icon 

Suited range 

 

< 1300 μS/cm 

 

1300 - 

1500 μS/cm 

 

   > 1500 µS/cm 

 

Measure

 

Rinse conductivity electrode in pure water, allow it to dry, and submerge it in the sample solution. Stir the solution 
briefly and allow it to stay in the sample solution until a stable reading is reached and stability

  

 

icon appears on LCD, then get the reading value which is the conductivity value of the solution.

 

 

During the process of calibration and measurement, the meter has self-diagnosis functions

 

 

Display Icons 

Contents 

Checking 

 

Wrong conductivity calibration solution or 
the  meter  recognition  of  calibration 
solution out of range 

1.Check whether conductivity calibration  

solution is correct. 

2.Check whether the meter connects the  

electrode well. 

3.Check whether the electrode is damaged. 

 

Press 

    

      key when measuring value is 

not stable during calibration. 

 
Press 

         

key

 

when     

   

 icon appears 

 

During  calibration,  the  measuring  value  is 
not stable for ≥3min.  

1.Shake  the  electrode  to  eliminate  bubbles  in 

electrode head. 

2.Replace with new pH electrode. 

 

Factory default setting

 

For factory default setting, please refer to parameter P2.8. All calibration data is deleted and the meter restores to the 
theory value. Some functions restore to the original value (refer to appendix -1). When calibration or measurement fails, 
please restore the meter to factory default setting and then perform re-calibration or measurement. Please note once 
set the factory default, all the data deleted will not be retrievable.

 

 

Содержание COND 7 Plus

Страница 1: ...1 7 Series Version 2 0 7 Series pH Conductivity Portable Meter Manual pH 7 DHS Portable pH Meter COND 7 Portable Cond Meter PC 7 DHS Portable pH Cond Meter...

Страница 2: ...ion consideration Standard buffer solution 9 Three point calibration 9 Calibration Intervals 9 pH Meter Calibration 10 Customer calibration 11 example of 1 60pH and 6 50pH calibration solution 11 Meas...

Страница 3: ...s between TDS and conductivity 18 Setup Menu 18 Main menu 18 Submenu 18 Submenu of pH parameter setting pH7 DHS PC DHS 20 Submenu of conductivity parameter setting COND7 PC7 DHS 21 Submenu of TDS para...

Страница 4: ...ower off and low voltage display The meter s digital filter improves measurement speed and accuracy There is reading stability display The package includes portable case meter electrode standard solut...

Страница 5: ...9 9 S cm 200 1999 S cm 2 00 19 99 mS cm 20 0 199 9 mS cm C0ND 7 PC 7 Resolution 0 1 1 S cm 0 01 0 1 mS cm Accuracy 2 0 FS Temperature compensation 0 80 manual or automatic Electrode constant 0 1 1 10...

Страница 6: ...eter Note In the calibration mode or the parameter set up mode press for return to the measurement mode then press to turn off the meter TDS 1 Parameter mode icons 2 Measurement reading 3 Prompts of s...

Страница 7: ...In the measurement mode press this key to enter in the parameter set up main menu In the calibration mode press this key to confirm calibration In the parameter set up mode press this key to select pr...

Страница 8: ...de in laboratory in ideal conditions using a XS DHS meter Calibration data like date slope model and batch number are memorized inside the electrode Moving this electrode on another XS meter DHS compa...

Страница 9: ...solution then select other standard solution to perform the second and the third point calibration the meter is able to recognize the buffer solutions in analysis During the calibration process the in...

Страница 10: ...er solution Stir the solution briefly and allow it to stay in the buffer solution until a stable reading is reached The meter s display will show scanning and locking process of calibration buffer sol...

Страница 11: ...to confirm it Then the main value blinks Follow the above procedures to adjust the main value and calibrate the meter Rinse pH electrode in pure water allow it to dry and submerge it in pH 6 50 buffe...

Страница 12: ...ange out of range 0 00 pH or 14 00pH 1 Check whether the electrode is suspended 2 Check whether the meter connects the electrode well 3 Check whether the electrode is damaged Factory default setting F...

Страница 13: ...shed with pure water and dried Do not clean the glass bulb with a tissue for it will affect the stability of the electrode potential and increase the response time The electrode should be thoroughly c...

Страница 14: ...tion for 6 hours Conductivity Conductivity cell Two ring cells with built in temperature sensor can be used on this instrument The conductivity cell uses a BNC connector while the temperature probe us...

Страница 15: ...form calibration from low to high concentration to avoid standard solution of low concentration being contaminated Reference temperature Reference temperature of factory default is 25 C Other referenc...

Страница 16: ...of calibration at 147 S cm Select CUS in the setup menu P2 2 the meter enters into customer defined calibration mode When press LCD shows blinking CUS at the top right indicating that the meter enter...

Страница 17: ...en get the reading value which is the conductivity value of the solution During the process of calibration and measurement the meter has self diagnosis functions Display Icons Contents Checking Wrong...

Страница 18: ...ter can only be calibrated in Conductivity mode and not TDS mode After calibration of conductivity the meter can switch from conductivity to TDS Adjust TDS conversion factor per parameter P3 1 accordi...

Страница 19: ...ard solution P2 5 Select reference temp P2 6 Adjust temperature compensation coefficient P2 8 Restore to factory setup P2 9 Temperature calibration P6 1 Select temperature unit P6 6 Parameters selecti...

Страница 20: ...press directional keys Yes blinks press key to confirm the meter returns to the measurement mode No Do not restore Yes Restore to factory setting 2 Press key to return to the measurement mode Set read...

Страница 21: ...perature value 15 0 30 0 press key to confirm After confirm parameter press key to return to the measurement mode Adjust temperature compensation coefficient 0 00 9 99 Press key 2 00 blinks press dire...

Страница 22: ...ress key to confirm When parameter is confirmed press key to return to the measurement mode Parameters selection Yes No for each parameters only PC7 Press key in the top left pH parameter is selected...

Страница 23: ...compensation coefficient 0 00 9 99 1 91 P2 8 Restore to factory default setting No Yes No P2 9 Temperature calibration Calibration range 5 Factory default value P3 0 TDS P3 1 Adjust TDS factor fact 0...

Страница 24: ...Calibration Temperature Calibration P2 0 Conductivity P2 1 Cell Constant Cell P2 2 Calibration solution Calibration solution P2 5 Reference temperature Reference temperature P2 6 Temperature compensat...

Отзывы: