background image

06

 

  7.2 SPEED SETTING

 

             

---Highest Speed          --Normal Speed            --Low Speed

 

 

 

 

 

 

 

 

 

 

 

 

 

7.3 WEIGHING UNIT SETTING   

  Under the weighing state, press 

“CONVERSION” to choose the unit. 

  There are 4 units :g/ct/oz/mg. 

   

  7. 4 COUNTING FUNCTION 

 

There are counting functions in the balance, And please keep the products 

 

have the same weight, and the minimum weight must be ≥0.5mg. 

 

 

 

 

 

 

 

 

 

 

 

1.Choose the sample quantity which you will put on the pan 

2.Touch the weighing value you want weighing,here is the sample weight,after   

    the weighing stable touch again, the result will check and store. 

3.Touch here and weighing the total weight, and touch again check the weight. 

4.The quantity will be show here. 

05

 

2.0 SPECIFICATIONS

 

Model 

BL224 Touch 

Range 

0-220g 

Division 

0.1mg 

Repeatability 

0.2mg 

Linearity 

0.3mg 

Resp. Time 

≤ 4 Sec. 

Pan Size 

ø80mm 

Interface 

RS-232 bi-directional 

Overall 

Dimensions 

340 x 215 x 350mm 

Operating 

temperature 

+10°C to 40°C 

Power Supply 

Output voltage 5 VDC, 600mA through 

External Power Adapter as standard (Input Voltage220 V) 

Net Weight 

8.5 kg 

 

3.0 UNPACKING THE BALANCE 

Remove the balance from the packing by carefully lifting it out of the box. 

Inside the box you will find everything needed to start using the balance 

 

4.0 CALIBRATION THE BALANCE 

Press”CAL”  key,  and  it  will  show  and  shining:  CAL-XXX.XXXX,and  the 
balance begin internal calibration. 

You also can choose manual calibration with standard weight, and there will 

have the notes tell you step by step: 

 

 

 

 

 

 

 

 