Start
Place the scale, and using the bubble on the backside for level It.
Press On / Off to turn on the scale. Following a countdown, marked also by the beep, you will see the
software version, the capacity of the scale, and then you enter the measurement mode.
Wait at least 30 minutes to warm up before the first use.
Date and Time settings
Switch on the balance. Keep pressed
for 3 seconds for date and time settings menu: “
rtc-set”
.
First value requested is the year: the display shows “
Year **”:
Use the following buttons to modify the blinking value:
set the value,
scroll to the next value
confirm.
Successively are required:
1.
Month
Month
2.
Day
Day
3.
Hour
Hour
4.
Minute
Min
User Calibration
Each balance is calibrated in the laboratory by technicians using certified sample masses.
Provided with all models there is a sample mass useful to test the functionality of the instrument
.
Calibration procedure
:
o
Turn on the balance and wait at least 30 min. of warmup
o
Keep pressed
button for 3 seconds,
the display shows first “
Cal 0
”, “
-------
“, and then it shows the sample mass [g]
requested for the calibration. *
o
Place the sample mass on the plate.
o
Display first shows “-------“, than the weight measurement.
Calibration is complete.
* With the
button is possible to change this value, choosing one of the allowed for that model (Tab 1)