background image

 

 

Waterproof tester series 

6

 

CALIBRATION PROCEDURE FOR CONDUCTIVITY (Cond1) 

1.

 

Power ON the meter by pressing the   Key. 

2.

 

Rinse the probe with distilled water.  

3.

 

Immerse the probe in the calibration solution (1413µS or 12.88 mS), wait for stability 

4.

 

Press 

CAL

 key. 

5.

 

The meter will start calibration procedure and will recognise automatically the standard 
used. 

6.

 

When stable press 

 to confirm and complete the calibration. 

7.

 

The standard value will flash for 3 times and then the meter will go into the measurement 
mode. 

8.

 

If the 2

nd

 point calibration is required then rinse the probe with distilled water and 

immerse in the second standard solution (1413µS or 12.88 mS), wait for stability 

9.

 

Repeat the points 4 to 7. 

10.

 

The calibration process is completed and the meter is ready to use. 

Note: Anytime press   key to abort and exit from calibration procedure.  

 
 

 Reading based on theoretical cell value C=1 

 Standard solution 

 

CALIBRATION PROCEDURE FOR TDS (Cond1) 

When the meter is set to read 

TDS

 then the calibration is done on TDS with 1 or 2 points. 

The calibration procedure for TDS is same as for Conductivity. 

CALIBRATION PROCEDURE FOR pH (pH1) 

1.

 

Power ON the meter by pressing the   Key. 

2.

 

Rinse the electrode with distilled water. 

3.

 

Immerse the electrode in the 1

st

 buffer solution pH7.0 and wait for stability 

4.

 

Press 

CAL

 key. 

5.

 

The meter will start calibration procedure and will recognise automatically the standard 
used. 

6.

 

When stable press 

 to confirm and complete the calibration. 

7.

 

The standard value will flash for 3 times and then the meter will ask for next point for 
calibration. If only 1 point calibration is required then press   to finish and exit. 

8.

 

If the 2

nd

 point calibration is required then rinse the electrode with distilled water and 

immerse in pH4.0 or pH10.0, wait for stability 

9.

 

Repeat the points 4 to 7. 

10.

 

The calibration process is completed and the meter is ready to use. 

 
Note: Anytime press   key to abort and exit from calibration procedure. 
Note2: when the first point calibration is confirmed (point 7) if the sensor is not removed from the 
buffer solution, the instrument may give wrong buffer error. 

 

pH 

6.9 

pH 

COND 

1460 

µS 

14 13 

pH 

7.2 

pH 

7.0 

COND 

14 13 

µS 

pH 

7.0 

pH 

COND 

1426 

µS 

COND 

14 13 

µS 

pH 

7.0 

pH 

COND 

1460 

µS 

14 13 

Содержание 1-5 Series

Страница 1: ...Version 1 0 _ March 2016 Waterproof tester series Version 1 0 _ March 2016 INSTRUCTION MANUAL TESTER SERIES 1 5...

Страница 2: ...WER OFF 5 SETUP MENU for pH1 Cond1 5 MEASUREMENT 5 CALIBRATION PROCEDURE FOR CONDUCTIVITY Cond1 6 CALIBRATION PROCEDURE FOR TDS Cond1 6 CALIBRATION PROCEDURE FOR pH pH1 6 INSTRUCTIONS FOR pH5 Cond5 PC...

Страница 3: ...tion Accuracy Automatic scale 1 f s Automatic scale 1 f s MultiPoint Calibration 1 2 1 3 Temperature compensation 1 9 C 0 00 4 00 C Reference temperature 25 C 20 25 C TDS Range 0 1ppm 10ppt 0 1ppm 10p...

Страница 4: ...ith fixed sensor and only two keys for all functions Tester series 5 is advanced version with replaceable sensor multicolour backlight display and 3 key for all functions SAFETY INSTRUCTIONS Read this...

Страница 5: ...Cond TDS SAL During Setup Press to scroll in the menu or increase the value of the selected parameter CAL During measurement press to start the calibration of the selected parameter Press to confirm t...

Страница 6: ...00 pH 1413 uS High 10 01 pH 12 88 mS First point for pH calibration is always 7 00 pH POWER SUPPLY This tester series works with 4 x AAA Alkaline or NiMH rechargeable batteries Make sure the batterie...

Страница 7: ...then the display will show TDS Fct flashing press to change this factor and then press to confirm Only for Cond1 5 The display will show rSt RESET nO flashing 6 Press and select YES if a reset of the...

Страница 8: ...TDS with 1 or 2 points The calibration procedure for TDS is same as for Conductivity CALIBRATION PROCEDURE FOR pH pH1 1 Power ON the meter by pressing the Key 2 Rinse the electrode with distilled wat...

Страница 9: ...cted parameter 5 The value of selected parameter will start flashing 6 Press to change the value and then press to confirm 7 The value stops flashing 8 Press to select other parameters or press ESC to...

Страница 10: ...pH5 PC5 1 Power ON the meter by pressing the Key 2 Rinse the electrode with distilled water 3 Immerse the electrode in the 1st buffer solution pH7 00 and wait for stability 4 Press CAL key blue backl...

Страница 11: ...istiller water or sample before use For short time storage use distiller water For long time store dry pH1 pH5 pH5 Food ORP5 PC5 Rinse the probe with distiller water or sample before use Store the pro...

Страница 12: ...bbles in glass bulb 2 Check whether pH buffer solution is correct 3 Replace with new pH electrode Electrode slope out of range 85 or 110 pH measuring range out of range 0 00 pH or 14 00pH 1 Series 2 0...

Отзывы: