4)  Should the bottom measuring tip touch the block first, you can use the torx screwdriver 

and make a new adjustment of your lower mess point. Loosen the screw from the shaft 
holder and move the shaft. Then fix the screw again. 

 

 
 
 
 

 
 
 
 
 

 

5)  Set the digital dial gauge to 10.0000 mm and make some measurements again. The 

value must be 10.0000 mm. 

 

6)  To  know  how  save  your  repeatability  measurement  is,  you  can  form  the  arithmetic 

mean value with this formula: 

 

 

𝑥 =

𝑥

1

+𝑥

2

+⋯𝑥

𝑛

n

 

 

7)  If you have big deviations, please check the situation/quality of the special measuring 

tips. It is important that you use only the Xonox measuring tips, no ruby or similar. 

 

8)  To control the mechanical axis, demount both measuring tips and put the measuring 

sleeve on the shaft. 

 

 

 

 

Содержание CT 200

Страница 1: ...ransducer and tighten the chuck 2 Switch on the digital dial gauge and hold the SET ZERO button for two seconds Now in the display appears 10 0000 mm and P flashes Pressing the Set button to set this value 10 0000 mm 3 Operate the left switch the measuring tips touch the 10mm gauge block It is important that the upper measuring tip first touch the gauge block Then press the SET button on the remot...

Страница 2: ... mm and make some measurements again The value must be 10 0000 mm 6 To know how save your repeatability measurement is you can form the arithmetic mean value with this formula 𝑥 𝑥1 𝑥2 𝑥𝑛 n 7 If you have big deviations please check the situation quality of the special measuring tips It is important that you use only the Xonox measuring tips no ruby or similar 8 To control the mechanical axis demoun...

Страница 3: ...Clamp the measuring sleeve in the chuck and let the upper and lower shaft glide into the measuring sleeve For this action the upper boom can be held by hand and the lower boom can be operated with the pneumatic left switch When you see that the lower shaft entering into the measuring sleeve lead the upper shaft by hand in the measuring sleeve 12 When the alignment is centered both shafts slide int...

Страница 4: ...iBall Ø 25 400 Tolerance for the measurements are 4 µm 14 It is absolutely forbidden to set the dial gauge to zero when the measuring tips stay in contact In additional it is very important that the alignment of the CT don t be checked in this way 15 Use always the gauge block to set the digital dial gauge to zero ...

Отзывы: