5.
Function and measurement
5.1.
Function of the measuring gauge
Illustration 9: Switch of the Mitutoyo 543-564D
1.
On-off button of the measuring gauge
2.
Adjusting the preset of 10mm:
-
Press and hold the [SET/ZERO] key for long time
Change between INC- and Preset system
-
Press the [SET/ZERO] key quickly (Preset-system)
adjustment
-
Press and hold the [SET/ZERO] key for long time
(Preset-system)
Leading sign is accentuated
Press Mode to change the leading sign
-
Press the [SET/ZERO] key quickly (Preset-system)
Moves to next digit
Adjust the preset of 10mm
-
Press the [SET/ZERO] key quickly (Preset-system)
Run over the Preset menu until nothing is
accentuated anymore
-
Press SET on the radio control
Preset value is supposed automatically
5.2.
Calibration
Illustration 10: Operation of calibration with gauge block
10mm
1.
Inlay the gauge block in the respective
holder
2.
Clamp the gauge block
3.
Upper probe DOWN
4.
Lower probe UP
5.
Reset the value 10mm of the measuring
gauge
6.
Lower probe DOWN
7.
Upper probe UP
Depending on the operation accuracy, the
calibration should be regularly repeated.
1
1
2
4; 6
3; 7