WALKAIDE II USER MANUAL
29
MNL-0023
Q6. What should I do if there is sporadic strong electrical
stimulation?
• Wet the skin with some water or the connect gel so as to increase the conductivity
between the electrode and the skin.
• Check whether the skin in the area covered by the electrode is red or has a wound.
• Check whether the stimulator has been fastened well to the leg or the electrode has
been placed on the current position.
Q7. Why can
’t I feel the stimulation when there should be
stimulation output?
Normally it is because the cuff position has been changed or the Gait mode has been
changed. Please readjust the stimulator placement or reset the parameters of Gait
mode.
Q8. Can I use oil or lotion on my legs?
No, please make sure the skin is clean before using the stimulator and wet the skin
with some water or the connect gel so as to increase the conductivity between the
electrode and the skin.