User Manual
Bobcat-640-GigE&CL Camera
Company confidential.
This document is the property of Xenics. It may not be reproduced – completely or partially – or passed to a third party
without written permission from Xenics.
Xenics nv
T +32 16 38 99 00
Doc Ref: ENG-2012-UMN020
Ambachtenlaan 44
F +32 16 38 99 01
Issue: R000.12
BE-3001 Leuven
Belgium
www.xenics.com
Date: 29/10/2013
XF-104_02/20-01-2012
Page 16 of 24
4. Electrical Interface
4.1. General Overview Connectors and Specifications
Interface
Connector
Specification
Camera Protocol
Bobcat-640-GigE
Input power (12V DC)
Hirose HR10-7R-4SA(73)
12V ±10%
Trigger
SMA
Trigger in:
V
IN,L
= 0.8V Max.
V
IN,H
= 2V Min.
V
IN,MAX
= 30V
Internal Pull-down: R = 10k
Ω
Trigger out:
V
HIGH
= 3.3V ±10%
V
LOW
= 0V
Ethernet
RJ45 connector
GigE standard
GigE Vision
Bobcat -640-CL
Input power (12V DC)
Hirose HR10-7R-4P(73)
12V ±10%
Trigger
SMA
Trigger in:
V
IN,L
= 0.8V Max.
V
IN,H
= 2V Min.
V
IN,MAX
= 30V
Internal Pull-down: R = 10k
Ω
Trigger out:
V
HIGH
= 3.3V ±10%
V
LOW
= 0V
Mini-camera link
CONN SDR 26POS
VERT RECEPT
Serial control: 115200 baud, 8n1
Levels: RS-644
XSP Protocol
(see
Image acquisition: CL
CL Base protocol/
1 TAP for image
acquisition
Table 4-1 Electrical interface specifications for Bobcat-640-GigE, CL Interface