OPERation Status Register
M370078-01
5-37
5
Table 5-12
OPERation SHUTdown Status Register
Bit
Bit Weight
Bit Name
Description
0
1
PROTection
Reflects the summary of the PROTection sub-
register.
1
2
INTerlock
The power supply is shut down by INTerlock signal.
2
4
Not Used
Not Used
3
8
External Shutdown
The power supply is shut down by External
Shutdown signal.
Table 5-13
OPERation SHUTdown PROTection Status Register
Bit
Bit Weight
Bit Name
Description
0
1
OVP
Over Voltage protection has tripped.
1
2
UVP
Under Voltage protection has tripped.
2
4
OCP
Over Current protection has tripped.
3
8
Not Used
Not Used
4
16
AC Off
AC failure protection has tripped.
5
32
OTP
Over Temperature protection has tripped.
6
64
Not Used
Not Used
7
128
Foldback
Foldback protection has tripped.
8
256
Fan
Fan failure protection has tripped.
Содержание Watt Series Programmable DC Power Supply XG 850
Страница 2: ......
Страница 3: ...XG 850 Watt Series Programmable DC Power Supply Operating Manual...
Страница 6: ......
Страница 18: ...xvi...
Страница 30: ...1 8...
Страница 52: ...Local Operation 3 8 M370078 01 Figure 3 3 Front Panel Menu System...
Страница 82: ...Local Operation 3 38 M370078 01 The SCPI command s for these instructions are SYSTem channel s REMote STATe...
Страница 86: ...Local Operation 3 42 M370078 01 Figure 3 5 Split Supply Operation V V Common Power Supply Power Supply Load...
Страница 126: ...Analog Programming APG and Isolated Analog Programming ISOL 4 34 M370078 01...
Страница 198: ...5 72...
Страница 262: ...A 36...
Страница 284: ...IX 4...
Страница 285: ......