Package Contents
1
RadTarge II
1
MicroUSB Charging Cable
1
User Guide
1
Certificate of Quality
Technical Parameters
Product Parameters
Detector
Measurement Range
Measurement Range
Dose Rate Linearity
Energy Range
Energy Response
Alarm Threshold Value
Alert Options
Alarm Response
Data Storage
Power / Data Transfer
Power Supply
Battery Life
Dimensions
YSO Scinti SiPM
Dose Rate: 0.01μSv/h ~ 1mSv/h
Integrated Dose: 0.01
μSv ~ 100
Sv
≤ 10% up to 1
mSv/h
20
keV ~ 3
MeV
≤ ± 20% (
∝
Cs-137)
1
μSv/h ~ 1
mSv/h
Audible (80 dB at a distance
of 12 in / 30 cm), Visual (LED &
Display) & Vibrating
≤ 2 sec
Accumulated Dose
Accumulated Dose per
H
our
Via MicroUSB and
PC
-
Based Software (RadSuite-Dose)
Rechargeable
Lithium-Ion
Battery
200 hours
2.7 x 1.8 x 0.7 in / 69 x 46 x 17 mm
Compliance
EMI/EMC
Complies with IEC 61526
FCC
2AC7P-110
Part
Model / Spec
Circuit Board
FR-4
Scintillation Crystal
YSO / SiPM 60035
Plastic Shell
PC-ABS
Single Chip
MSP430F437
Possible Problems
Solutions
Unable to Turn On
Check whether the power
button was pressed for longer
than 2 seconds. Make sure
battery has been fully charged.
Automatic Shutdown
Make sure battery has been
fully charged.
Insufficient Power
Indicator - ON
Make sure the battery has
completed charging.
Charging Indicator
Light Is Not Blinking
Ensure MicroUSB is
connected properly. If problem
persists, choose another port
to plug in USB.
Continuous Alarm
Ringing
A radiation source is nearby.
Check threshold settings.
Alarm Is Not Ringing
Near Radiation Source
Check threshold settings.
*If the problem listed above can not be solved, please
contact X-Z LAB or your Account Administrator.
Electronic Personal Dosimeter
USER GUIDE
X-Z LAB, Inc.
Bishop Ranch 6
2440 Camino Ramon
Suite
#
264
San Ramon, CA 94583
United States
Phone
(925) 359-6908
Fax
(925) 380-6784
Email [email protected]
www.x-zlab.com
Real Time Setting
To enter setup mode, press the settings button 5
times. When ‘RTC’ is displayed, press power button
to enter setup.
Press the settings button to choose Year, month, day
and time. Press power button to apply changes. Press
settings button until ‘DONE’ is displayed to complete
process.
Press settings button to return to the main interface.