C i 5 1 / C i 5 2 S P E C T R O P H O T O M E T E R
16
Black Trap Cleaning Procedure
1.
Take apart the two sections by compressing the two locking tabs on both sides with your
fingers and separating.
2.
Clean with clean, dry air, or wipe clean with a lint-free cloth to remove any dust or
contamination.
3.
After cleaning, align the trap opening in the top section over the cone in the bottom section of
the reference. Assemble the two sections.
Note: The tabs will not lock correctly if assembled in the opposite direction.
White ceramic plaque with
protective cap
Trap cone
Locking tabs
Trap opening