background image

     

      

 

 

 

 

 

 

 

 

 

SIDE 2

 

X-Cite

®

 200DC: Quick Start Guide

 

            

Warning 

– Refer to User Manual 035-00446R for all safety precautions prior to use.

 

 

                                                                                              

Lumen Dynamics Group Inc. 

 

Tel: 1-905-821-2600 

2260 Argentia Rd. 

Toll Free: 1-800-668-8752 

Mississauga, Ont.  

Fax: 1-905-821-2055 

Canada, L5N 6H7 

www.ldgi-xcite.com 

 

 

 

 

Installing the Lamp Module 

1.  Be sure the AC power cord is disconnected from the unit. 
2.  Remove the screw from the lamp access side panel using the 3mm hex 

tool. 

3.  Carefully remove the lamp module from its container, holding only the 

ceramic component and glass rim 

4. 

As illustrated below, position the lamp facing towards the front of the unit 
with the power leads facing towards you.  

 

Caution: 

Do not 

touch the bulb’s glass envelope or the inside surface of 

the reflector. Skin oils can cause the lamp module to fail prematurely. 

 

5.  Make sure the middle of the lamp is in position to fit into the spring 

clamp. Slide the lamp until it snaps into the spring clamp. The leading 
edge of the reflector should fit snugly into the lamp holder recess. 

 

 

 

6.  Locate the 4-pin Intelli-Lamp sensor connector at the rear of the lamp module 

and connect it to its mate located on the top of the lamp-housing wall. 

       Tip:

 the Intelli-Lamp connector will only attach in the correct orientation. If 

       you are having difficulty attaching the connector, try rotating it by 180º. 

 

7.  Locate the Power connector with two leads and connect it to its mate located 

on the bottom of the lamp-housing wall. 

        

Tip:

 the Lamp Power connector will only attach in the correct orientation. If  

        you are having difficulty inserting the connector, try rotating it by 180º. 

 

8.  It is very important to ensure that the lamp anode cooling fin (bar) and 

lamp power lead, at the front of the lamp are centered within the lamp 
holder cutout.  Rotate the lamp as required. 

9.  Replace the lamp housing side panel and tighten the fastening screw. 
10.  Replace the hex tool in the clips underneath the system for future use. 

Note:

 if the lamp housing panel is not secured completely, the lamp will not 

strike and the “

bulb

” message will display when power is turned on to the unit. 

 

 

Rim 

*Center 
lamp 
power 
lead as 
shown