X-Cite
®
120PC Q: Quick Start Guide
SIDE 2
Lumen Dynamics Group Inc.
Tel: 1-905-821-2600
2260 Argentia Rd.
Toll Free: 1-800-668-8752
Mississauga, Ont.
Fax: 1-905-821-2055
Canada, L5N 6H7
www.ldgi-xcite.com
Installing the Lamp Module
1.
Be sure the AC power cord is disconnected from the unit.
2.
Remove the screw from the lamp access side panel using the 3mm hex tool
provided and remove the panel from the unit cover.
3.
Open the Lamp bracket arm by pulling towards you and to the right.
4.
Carefully remove the lamp module from its container, holding only the
ceramic components.
Caution:
Do not touch the bulb’s glass envelope or the inside surface of the
reflector. Skin oils can cause the lamp module to fail prematurely.
5.
With the lamp facing towards the rear of the unit and “This Side Out” label
facing out, position the lamp so that the two leading edges of the lamp
ceramic mount slide into the groove of the lamp bracket. The middle of the
lamp should be in position to fit into the spring clamp.
6.
Slide the lamp all the way in so that both leading edges of the lamp ceramic
mount are in the upper and lower grooves of the lamp bracket. The middle
of the lamp will snap into the spring clamp.
7.
Close the Lamp bracket arm by pushing it in and to the left.
8.
Insert the 4-pin Intelli-Lamp connector into it’s mating receptacle on the lamp
housing wall as illustrated below.
Tip:
the Intelli-lamp connector will only attach in the correct orientation. If you
are having difficulty attaching the connector, try rotating it by 180º.
9.
Attach the 2-pin Lamp Power connector to its mating receptacle on the lamp
housing wall as illustrated below.
Tip:
the Lamp Power connector will only attach in the correct orientation. If you
are having difficulty inserting the connector, try rotating it by 180º.
10. Insert the Intelli-Lamp lead harness and the 2 Lamp Power leads into the
alignment grooves on the lamp housing as illustrated below.
11. Replace the lamp access side panel and tighten the fastening screw.
Note:
if the lamp is not properly installed, the “
bulb
” message will display and
the lamp will not strike.