Specifications: 

Measurement Ranges (Probes up 3 or 4 Tesla on request): .................... 2 mTesla, 20mTesla, 200 mTesla,  2Tesla 

(depending of the selected probe with 3 or 4 ranges) 

20Gauss, 200Gauss, 2 kGauss, 20 kGauss 

 

1,591kA/m; 15,91kA/m; 159,1KA/m; 1,591MA/m 

 

 15,91A/cm; 159,1A/cm; 1,591kA/cm; 15,91kA/cm 

 

20 Oersted; 200 Oersted; 2kOersted; 20 kOersted 

Auto ranging .......................................................  The Tesla / Gauss Meter switch automatically in the optimal range    

 

Display and switchable units : ......................................................................................................... 3 ½ digit display  . 

                           

1,999 mT, 19,99 mT, 199,9 mT, 1999 mTesla 

                           

 19,99 Gauss, 199,9 Gauss, 1,999 kGauss, 19,99 KGauss 

 

1,591kA/m; 15,91kA/m; 159,1KA/m; 1,591MA/m 

 

 15,91A/cm; 159,1A/cm; 1,591kA/cm; 15,91kA/cm 

 

19,99 Oersted; 199,9 Oersted; 1,999kOersted; 19,99 kOersted 

 

Accuracy: 

Device  ..................................................................................................................................... 0,2% FSR  ± 1 Digit  

Probe ...................................................................................................................... 0,3% FSR (DC)  / 2% FSR(AC)  

Long time stability: ................................................................................................................................ 0,1% per year  

Reproducibility: ............................................................................................................... 0,1% FSR (Units and probe)

 

Operating temperature: ..................................................................................................................... 10° C to + 45° C 

Storing temperature: ........................................................................................................................ -20° C to + 55° C 

Temperature coefficient: ...................................................................................... 0,01% F.S. per °C (Unit and probe) 

Functions: .................................................................................................. DC, negative and positive peak value AC,  

 

 

Power supply:  

Battery: ..................................................................... 3 x1,5 volt  AA Batteries (for approximately  70 hour operating) 

USB  ................................................................................  with external USB power supply or USB connection to PC 

 

 

Dimensions and weight:  

Dimension: ................................................................................................................. Unit 165mm x 78 mm x 34 mm 

Weight: ....................................................................................................................... Unit 255 Gramm (incl. Batterie) 

                           

Transvers probe 43 Gramm 

                           

Axial probe 55 Gramm 

Output and interface:  

Analog output: ............................................................................................ ±800 mV F.S. up to 10 KHz not corrected  

 

Connection by 2,5 mm mono connector 

Interface: ....................................................................................................................................................... USB 1.1 

Software ......................... Windows based Software for displaying and logging the measurement values at computer 

Displaying and data logging Software: 

The  software  enclosed  free  of  charge  shows  the  displayed  value  of  the 

device on the PC screen and offers the possibility of saving the readings 

in the selectable interval between 0,2 seconds and 50 seconds to the PC 

hard disk.  

 

As data format its possible to select between the  Excel compatible CVS 

or ASCII format. 

 

Remote control:  

The  KOSHAVA  5  can  be  remote  controlled  by 

Computer with the USB Interface 

 

The Magnetometer KOSHAVA 5 gets the power 

from  the  USB  interface,  if  the  unit  is  connected 

with the computer.

 

Amendments reserved.

 

Содержание KOSHAVA 5

Страница 1: ... com Heppstrasse 30 D 80995 Munich Phone 49 89 313 30 07 Fax 49 89 314 67 06 Copyright 2015 WUNTRONIC GmbH All Rights Reserved Magnetometer KOSHAVA 5 User Manual Digital Tesla Meter Gauss Meter KOSHAVA 5 with switchable units Tesla Gauss kA m und A cm analog output and USB interface ...

Страница 2: ...4 English or German Zero field adjustment 4 Probe selection 5 Probe dimensions 4 Software 6 Specifications 6 Warranty 7 Limited warranty for probes 7 Transport damage 7 Rugged upholstered carrying suitcase Tesla Meter Gauss Meter Order number 1099255 1 meter probe cable USB cable to connect to PC User manual in English and German language Software for displaying and logging the measure values Cali...

Страница 3: ... or the interconnecting probe cable onto the plug of the device so tightly that the latch on the left and right side clicks in At the first start of the unit with a new probe the Tesla Meter reads out from the probe the individual calibration and lineariza tion table For the first use or to measure very small magnetic fields within the smallest range is recommended to run the zero field adjust men...

Страница 4: ...eld measurement Polarity DC or AC Field measurement 5 6 USB interface The device is equipped with a USB interface The soft ware enclosed free of charge shows the displayed value of the device on the PC screen and offers the possibility of saving the readings in the selectable interval between 0 5 seconds and 100 seconds to the PC hard disk Power from the USB interface The Magnetometer KOSHAVA 5 ge...

Страница 5: ... German Zero field adjustment With the zero field adjustment the environment fields becomes sup pressed Zero field adjustment with the optional zero field chamber If you put the probe top during the zero field adjustment into a zero field chamber the geomagnetic field is suppressed Zero field adjustment without zero field chamber If you run the zero field adjustments without the zero field chamber...

Страница 6: ... with specials size and ranges Please call factory Probe dimensions 5 80 11 30 50 0 8 63 82 10 80 63 80 Field direction active area 0 3x0 3 mm or 1x2 mm depending of the probe 35 45 0 8 10 79 11 16 30 6 Axial probes Transverse probes Dimensions in mm active area 0 3x0 3 mm or 1x2 mm depending of the probe 5 ...

Страница 7: ... Functions DC negative and positive peak value AC Power supply Battery 3 x1 5 volt AA Batteries for approximately 70 hour operating USB with external USB power supply or USB connection to PC Dimensions and weight Dimension Unit 165mm x 78 mm x 34 mm Weight Unit 255 Gramm incl Batterie Transvers probe 43 Gramm Axial probe 55 Gramm Output and interface Analog output 800 mV F S up to 10 KHz not corre...

Страница 8: ...e This war ranty is in lieu of all other warranties expressed or implied and no representative or person is au thorized to represent or assume for WUNTRONIC any liability in connection with the sale of our products other than set forth herein Warranty The Tesla Meter and probe should be tested as soon as it is received If it is damaged in any way a claim should be filed with the carrier A full rep...

Отзывы: