Manual for use of thermo-hygro-barometer COMMETER D4130
Instrument is designed for measurement of temperature, relative humidity, atmospheric
pressure and pressure tendency. It enables to display the dew point temperature and atmospheric
pressure corrected to the sea level. Measured values can be stored in programmable interval into
internal non-volatile memory and transfer anytime to a PC. Measured values are displayed on a dual
line LCD display. Temperature is measured by RTD sensor Ni1000/6180ppm. Instrument compares
all measured values (except pressure tendency) with two adjustable levels for each measured
quantity. Breaking the level is indicated by blinking the proper value on display and by audio
indication (switchable). Instrument is equipped with minimum and maximum memory and Hold
function. Minimum and maximum values and Hold value are possible to display on the LCD
anytime.
Technical parameters:
Parameters of measurement:
Temperature: Range of measurement: -10 to +60 °C
Resolution:
0.1
°C
Accuracy:
±0.4
°C
Humidity:
Range measurement: 5 to 95 %RH
Resolution:
0.1
%RH
Accuracy: ±2.5 % RH at the range of 5 to 95 %RH at 23°C
Dew point temperature (calculated from temperature and humidity):
Range: -50 to +60 °C
Resolution:
0.1
°C
Accuracy: ±0.5 °C at the range of 30 to 95 % RH
Atmospheric pressure: Range of measurement: 800 to 1100hPa
Resolution: 0.1 hPa
Accuracy: ± 2hPa at ambient temperature of 23°C
Measuring interval and display reading refresh: approximately 0.7 s in FAST mode
0.7 to 5 s in dynamic mode
Response time (air flow approximately 1 m/s):
temperature: t63 < 35 s, t90 < 60 s (temperature step 20 °C)
relative humidity: t63 < 30 s, t90 < 60 s (humidity step 30 %RH, constant temperature)
Interval of refresh of pressure tendency display: 0.5 h
Logging interval (in automatic mode): 10s to 24hours
Memory capacity:
for manual logging
1000 sets of all input quantities
for non cyclic automatic logging
total 4062 values
for cyclic automatic logging
total 3938 values
The above specified numbers are maximum possible. Capacity is slightly lower if logging
was interrupted by communication with computer, by stopping logging or switching OFF
the instrument
Communication with computer: RS232 (serial port)
Real time clock: adjustable from computer, integrated calendar including leap years
Real time clock error: < 200 ppm (i.e. 0,02 %, 17,28 s in 24 h)
Power: battery 9V or ac/dc adapter 12V with NiCd accumulator 9V
Average current consumption: 0.3 to 1.3 mA (depending on operation mode)
IP protection: IP20