WUNDER SA.BI.SRL                                            EN                                            USER MANUAL 

 

BABY02 (WU150)    REV.02 10/01/2017                                                                                                                    4 

The electronic scales BABY 02 model is scheduled for operation in the electromagnetic environment 

specified below. 

the customer and the user should ensure that it is used in that environment. 

Immunity test 

Compliance 

Guide to electromagnetic environment 

Electrostatic 

discharges (ESD) 

IEC/EN61000 - 4 -2 

6kV in contact 

8kV in air 

The floors should be made of wood, concrete or ceramic. If 

the  floors  are  covered  in  synthetic  material,  the  relative 

humidity should be maximum 30%. 

Rapid 

transients/burst 

IEC/EN61000 - 4 - 4 

2kV power supply 

 

The  power  supply  should  be  of  the  type  used  typically  in 

commercial or hospital environments.   

Surge 

IEC/EN61000 - 4 - 5 

1kV 

differential 

mode 

 

The  power  supply  should  be  of  the  type  used  typically  in 

commercial or hospital environments. 

Voltage dips, short 

interruption and 

voltage variation 

IEC/EN61000 - 4 - 11 

<5%UT for 0.5 cycle 

40%UT for 05 cycle 

70%UT for 25 cycle 

<5%UT for 5 sec 

The  power  supply  should  be  of  the  type  used  typically  in 

commercial or hospital environments. Note=UT is the value 

of the voltage of the feed. 

Power frequency 

Magnetic field 

IEC/EN61000 - 4 - 8 

3A/m 

 

    NOTE: UT is the value of the power supply voltage. 

  CAUTION 

The  medical  device  requires  particular  electromagnetic  compatibility  precautions  and  must  be  installed  and 

used according to the information provided in the accompanying documents. 

 

Manufacturer's guide and declaration - Electromagnetic emissions 

The electronic scales BABY 02 model can be used in the electromagnetic environment specified below. The 

customer or the user of the seat must ensure that it is used in compliance with the specifications below. 

Emissions test   

Compliance 

Guide to electromagnetic environment 

Radiated/ 

conducted / 

emissions   

IEC/EN61000-4-6 

3Vrms  150kHz  to 

80MHz(for 

appliances  that  are 

not life supporting) 

Equipment in RF communications and mobile laptops should ot 

be used near any part of the tooth, including cables, except 

when they meet the recommended separation distance 

calculated by the frequency applied to the transmitter. 

Содержание BABY02

Страница 1: ...sing the instrument 2 GENERAL RULES 2 3 SAFETY 3 4 SPECIFICATION 6 5 INDICATOR 6 6 INSTALLATION 7 7 ISTRUCTION FOR USE 8 8 PRINTER FUNCTIONS 11 9 ERROR MESSAGE 12 10 MAINTENANCE AND ASSISTANCE 13 11 CONFORMITY 13 12 IDENTIFICATION LABELS 13 13 SCRAPPING AND WASTE DISPOSAL 13 14 WARRANTY 14 15 INSTALLATION BABY HEIGHT ROD 15 16 INSTALL SECURITY BRACKETS 16 Manual_Baby02_V1 ...

Страница 2: ... simultaneously view Weight and weight difference calculation 2 GENERAL RULES Carefully read this manual before using the instrument as it supplies important indications concerning OPERATING SAFETY AND MAINTENANCE The descriptions and illustrations supplied in this manual are not binding WUNDER reserves the right to modify its products as deemed convenient in order to approve them without being co...

Страница 3: ...ord does not create obstruction hazards Unplug the instrument before cleaning it Do not place the instrument in water or other liquids Perform maintenance and subsequent metric verifications regularly 3 1 Manufacturer s guide and declaration electromagnetic immunity Guide and Statement of manufacturer Electromagnetic emissions The electronic scales BABY 02 model is scheduled for operation in the e...

Страница 4: ... UT for 25 cycle 5 UT for 5 sec The power supply should be of the type used typically in commercial or hospital environments Note UT is the value of the voltage of the feed Power frequency Magnetic field IEC EN61000 4 8 3A m NOTE UT is the value of the power supply voltage CAUTION The medical device requires particular electromagnetic compatibility precautions and must be installed and used accord...

Страница 5: ... AM and FM and TV transmitters cannot be made theoretically and with precision To establish an electromagnetic environment due to fixed RF transmitters it should consider the electromagnetic survey of the site If the field strength measured at the place where you use the instrument exceeds the applicable level of compliance of the above it should be placed under observation normal operation of the...

Страница 6: ...apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people 4 SPECIFICATION Model BABY 02 Capacity Max 3 6kg e 1 2g Max 6 15kg e 2 5g Max 10 20kg e 5 10g Approvazione OIML Class III Unit weight kg Display Double LCD 1 LCD 20mm weight with 5 active digits 2 LCD 12mm Dimensions 550 x 290 DISH Key functions ON ZERO OFF HOLD TARA PRIN...

Страница 7: ...Display Symbols Stable symbol To indicate that the weight is stable Minus weight Weight under zero Zero symbols Weight is at zero point Low battery Battery need to charge or replaced 6 INSTALLATION After removing the instruments from the packing check the integrity and right contents Put the scale far from source from of heat e vibration Put the scale on stable surface Indicator assembly is not as...

Страница 8: ...he TIME SETTING mode beginning with the top row with the flashing digit Press PRINT key after successful change to move to next step EX To input Dec 25 2016 00am Enter year Using TARE keys to decrease the value press HOLD key to move to next digit Press PRINT key after successful entry to move to next stage Enter date Using TARE keys to decrease the value press HOLD key to move to next digit Press...

Страница 9: ...hed accurately even they keep moving Once the HOLD key is pressed the weight reading will remain on the display after the item has been removed from the scale so the reading can easily be read Switch ON the scale using The diagnostic self checks is performed The scale is ready for weighing when the 0 0 kg displays on the screen Place the baby in the center of the tray gently Press When the triangl...

Страница 10: ...e remove the tare object from the tray and press key again Using memory function Milk difference This function allows automatically calculates the amount of Milk consumed After weighting the top row display will show the weight for example 3 000kg Press key to store the first weight the second display will show m1 And then press again to return to normal mode When the second weight be stored for e...

Страница 11: ... ASCII Serial RS232 Connecting with PC Start Hyper Terminal Start Hyper Terminal program from clicking Start Menu Programs Accessories Communication Hyper Terminal New Connection Description Give new connection a name then click OK Select Your COM Port Click Connect to select your COM port Usually there s only one option for select Then click OK Port Settings Click Bits per second to set up rate a...

Страница 12: ...clusive adaptor Locate adaptor plug on the rear side of indicator The battery should be recharged at least every 3 months regardless if it is used or not After a long period in storage e g over 3 months the battery should run a full cycle charge discharge to allow it to restore to full capacity 9 ERROR MESSAGE AC Jack connector ...

Страница 13: ...nder it 11 CONFORMITY WUNDER MODEL BABY02 ELECTRONIC SCALE SERIAL N We hereby certify that this instrument has been inspected and has successfully passed the functional test It complies with the following standards and directives EN 45501 EN60601 1 2 EN60601 1 93 42 EEC Medical Devices Directive 12 IDENTIFICATION LABELS 13 SCRAPPING AND WASTE DISPOSAL If set aside for a long period protect those p...

Страница 14: ...ries containing Lead Cd Cd batteries containing Cadmium Hg Hg Batteries containing Mercury ATTENTION Do not throw electric parts and used batteries away with household waste Dispose of the batteries by means of your closest collection centres 14 WARRANTY This certificate must be kept until the warranty has expired It must be presented together with the invoice tax receipt or delivery note providin...

Страница 15: ...e s home the customer must pay the fixed fee if however the appliance is repaired at an authorised Wunder Service Centre expenses and relative travel risk are borne by the user Wunder will not be held liable for damage of any nature caused directly or indirectly to persons animals or objects resulting from failure to comply with all the instructions indicated in this manual or anyway resulting fro...

Страница 16: ...MANUAL BABY02 WU150 REV 02 10 01 2017 16 16 INSTALL SECURITY BRACKETS FIG 1 1 LOOSEN THE PLATE AND REMOVE THE FEET fig 1 FIG 2 2 PLACE THE BAR AND SCREW THE ADJUSTABLE FEET fig 2 FIG 3 3 TURN THE SCALE AND MOUNT THE PLATE fig 3 ...

Отзывы: