wtw WTW Turb 430 IR Скачать руководство пользователя страница 8

Initial commissioning

Turb 430 IR/T

8

ba75507d04

11/2018

Initial commissioning

Switching on the meter

Press the 

<ON/OFF>

 key.

Setting the language

The English language is set on delivery. Set a different language as 
follows:

Setting the date and

time

The date and time are set in the menu, 

Configuration

 / 

System

 / 

Continue ...

 / 

Date/time

.

Operation

Inserting a cell

To be able to insert cells in the Turb 430 IR/T, the cell shaft has to be 
prepared to take in a cell.

i

 Insert standard

1000 FNU/NTU

i

 Press and hold <START>

i

 Align sample

Turbidity \ Calibration

The 

i

 indicates 

info texts, e.g. 
messages, notes or 
instructions

1

Open the 

Configuration

 menu with the 

<MENU>

 key.

2

Open the 

Configuration

 / 

System

 / 

Language

 menu with the 

<

> <

>

 and 

<START/ENTER>

 keys.

3

Select the required language with the 

<

> <

>

 keys and 

confirm with 

<START/ENTER>

.

4

Quit the menu with the 

<M>

 key.

1

Push the dust cover (1) upward. 
The cell shaft for 28 mm cells is open.

Содержание WTW Turb 430 IR

Страница 1: ...BEDIENUNGSANLEITUNG ba75507d04 11 2018 Turb 430 IR T TASCHENTURBIDIMETER...

Страница 2: ...kter Form die Beschreibung aller Grundfunktionen alle Hinweise f r den sicheren Betrieb und alle technischen Daten Eine ausf hrlichere Beschreibung mit Hinweisen f r besondere Anwendungen erhalten Sie...

Страница 3: ...a 14 Turbidity Turb 430 IR 16 Turbidity Turb 430 T 16 Note Part of the process of consequently improving our products is the continuos further development of instrument firmware All current data for t...

Страница 4: ...ting manual Safety instructions Safety instructions point out dangers Caution indicates instructions that must be followed precisely in order to avoid the possibility of slight injuries or damage to t...

Страница 5: ...ntains basic information on the operation of the Turb 430 IR T Operating modes Measurement The display indicates measurement data in the measured value display Calibration The display indicates a cali...

Страница 6: ...edures contain further submenus The selection is made with the keys The current selection is displayed in reverse video Menus The name of the menu is displayed at the upper edge of the frame Switch to...

Страница 7: ...TER Example Functions Functions are designated by the name of the function They are immediately carried out by confirming with START ENTER Example display the Calibration record function in the Turbid...

Страница 8: ...To be able to insert cells in the Turb 430 IR T the cell shaft has to be prepared to take in a cell i Insert standard 1000 FNU NTU i Press and hold START i Align sample Turbidity Calibration The i ind...

Страница 9: ...cell see below 1 Clean the cell 2 Insert the cell 3 Align the cell Press and hold the START ENTER key Slowly and in small steps rotate the cell by one complete rotation by 360 After each step wait fo...

Страница 10: ...lack light protection cap 1 Rinse out a clean cell with the sample to be measured Pour approximately 10 ml sample into the cell Close the cell and rotate it several times before throwing the sample aw...

Страница 11: ...free The only maintenance task is replacing the batteries or accumulator pack Caution Make sure the poles of the batteries are the right way round The signs on the batteries must correspond to the si...

Страница 12: ...ge the optical characteristics of the cell and falsify the measured value For this reason never use scratched cells 1 Open the battery compartment Unscrew the two screws 1 on the underside of the mete...

Страница 13: ...ries Charge the accumulator pack Cause Remedy Software error Operating condition undefined or EMC load unallowed Processor reset Press the START ENTER and PRT key simultaneously Cause Remedy Instrumen...

Страница 14: ...Operating time with battery operation Turb 430 IR approx 3000 measurements Turb 430 T approx 2000 measurements Accumulator pack optional 5 x 1 2 V nickel metal hydride NiMH type AAA Power pack Chargin...

Страница 15: ...digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment...

Страница 16: ...red value or 0 01 FNU NTU Reproducibility 0 5 of the measured value Response time 4 seconds Calibration Automatic 3 point calibration Measuring principle Nephelometric measurement according to US EPA...

Страница 17: ......

Страница 18: ...buildings factories and farms In more than 150 countries we have strong long standing relationships with customers who know us for our powerful combination of leading product brands and applications...

Отзывы: