Overview

1.1

Structure and function

Structure

Automatic sensor

recognition

The sensor electronics with the stored sensor data is in the connecting head. 
The data include, among other things, the sensor type and series number. 
With each calibration, the calibration data is written in the sensor and the 
calibration history is recorded. The data is recalled by the meter when the 
sensor is connected and is used for measurement and for measured value 
documentation. Storing the calibration data in the sensor ensures that the 
correct cell constant is automatically used if the sensor is operated with 
several meters.

The digital transmission technique guarantees the failure-free communi-
cation with the meter even with long connection cables. If the sensor firmware 
is enhanced by WTW

,

 it can be updated with the meter.

1

Voltage electrode (inside, 2x)

2

Current electrode (ring, 2x)

3

Temperature sensor in graphite enclosure

4

Shaft

5

Connecting head with active sensor electronics

6

Connection cable (TetraCon

®

 925/LV)

7

IDS plug (TetraCon

®

 925/LV-P)

6

4

1, 2

3

5

7

TetraCon 925/LV-P

®

TetraCon 925/LV

®

Содержание TetraCon 925/LV

Страница 1: ...OPERATING MANUAL ba77131e02 12 2015 TetraCon 925 LV P CONDUCTIVITY MEASURING CELL TetraCon 925 LV TetraCon 925 LV P...

Страница 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...

Страница 3: ...f application 6 2 Measurement Operation 6 2 1 Commissioning 6 2 2 Opening and closing the IDS plug connection TetraCon 925 LV P 7 3 Cleaning 8 4 What to do if 8 5 Technical data 9 5 1 General data 9 5...

Страница 4: ...Contents TetraCon 925 LV 4 ba77131e02 12 2015...

Страница 5: ...r measurement and for measured value documentation Storing the calibration data in the sensor ensures that the correct cell constant is automatically used if the sensor is operated with several meters...

Страница 6: ...re TetraCon 925 LV Connect the sensor to the meter The sensor is immediately ready to measure TetraCon 925 LV P Connect the sensor to a free IDS sensor plug in position of the multi parameter probe or...

Страница 7: ...the locking device step 1 Use your thumb and index finger to press the clips of the connector togeth er and pull the connector out of the plug step 2 Closing the plug connection Make sure that the plu...

Страница 8: ...cell as electronic waste 4 What to do if Contamination Cleaning procedure Lime sediments Immerse in acetic acid for 5 minutes vol ume share 10 Fat oil Clean with warm water that contains wash ing up...

Страница 9: ...0 k at 25 C 77 F Dimensions in mm 16 16 120 0 120 0 176 0 222 8 15 3 15 3 21 7 21 7 TetraCon 925 LV TetraCon 925 LV P Weights TetraCon 925 LV 60 g without cable TetraCon 925 LV P 65 g Materials Shaft...

Страница 10: ...ets the requirements according to article 3 3 of the directive 97 23 EC pressure equipment directive Measurement conditions Conductivity measuring range 1 S cm 2 S cm Temperature range 5 70 C 100 C 23...

Страница 11: ...199 9 200 1999 0 1 1 mS cm 2 00 19 99 20 0 199 9 200 2000 0 01 0 1 1 resistivity Ohm cm 0 50 19 99 20 0 199 9 200 1999 0 01 0 1 1 resistivity kOhm cm 2 00 19 99 20 0 199 9 200 1999 0 01 0 1 1 resistiv...

Страница 12: ...ection cable 10 m AS IDS 10 903 853 IDS connection cable 15 m AS IDS 15 903 854 IDS connection cable 20 m AS IDS 20 903 855 IDS connection cable 25 m AS IDS 25 903 856 IDS connection cable 40 m AS IDS...

Страница 13: ......

Страница 14: ...buildings factories and farms In more than 150 countries we have strong long standing relationships with customers who know us for our powerful combination of leading product brands and applications...

Отзывы: