Measuring / operation
SenTix
®
FET-D/B
30
Connecting the
sensor
SenTix FET-B (with BNC plug):
l
Connect the sensor to a pH meter. Plug the BNC plug in the corresponding
socket of the meter and lock it by turning it to the right. Connect the plug
of the temperature sensor to the corresponding socket of the meter.
SenTix FET-D (with DIN plug):
l
Connect the sensor to a pH meter. Plug the DIN plug and the plug of the
temperature sensor in the corresponding sockets of the meter.
Switching on the
sensor
l
Press the On/Off key to switch on the sensor. The measuring LED flashes.
Adjusting the zero
point
l
Place the sensor in a vessel with pH 7.00 buffer solution. When doing so,
the pH sensitive area and the junction must be completely submersed in
the solution. If necessary, slightly move the sensor in the solution to
remove air bubbles in the vicinity of the junction and pH sensitive area.
Plug of the temperature
sensor
BNC plug
Plug of the temperature
sensor
DIN plug