background image

Wear parts and accessories

SensoLyt

®

 ORP 900-P

8

ba75993d03

06/2017

The electrodes meet the requirements according to article 3(3) of the direc-
tive 97/23/EC ("pressure equipment directive").

7

Wear parts and accessories

Accuracy of the

IDS measuring

technique

Measured parameter

Accuracy (± 1 digit)

U [mV]

± 0.2

T [°C]

± 0.1

Pressure range at

temperature

Temperatureallowed overpressure

0 °C (32 °F)

1000 kPa (10 bar)

20 °C (68 °F)

1000 kPa (10 bar)

30 °C (86 °F)

500 kPa (5 bar)

40 °C (104 °F)

300 kPa (3 bar)

60 °C (140 °F)

100 kPa (1 bar)

Storage

With watering cap; filled with KCl 3 mol/L, Ag

+

 free

Disposal

Residual waste

Maintenance

equipment

Description

Model

Order no.

Reference electrolyte solution 250 ml 
to fill the watering cap
(KCl 3 mol/l, Ag

+

-free)

KCl-250

109 705

ORP buffer solution (250 ml)

RH 28

109 740

Connection cable

SensoLyt

®

 ORP

900-P - meter

Description

Model

Order no.

IDS connection cable, 1.5 m

AS/IDS-1.5

903 850

IDS connection cable, 3 m

AS/IDS-3

903 851

IDS connection cable, 6 m

AS/IDS-6

903 852

IDS connection cable, 10 m

AS/IDS-10

903 853

IDS connection cable, 15 m

AS/IDS-15

903 854

IDS connection cable, 20 m

AS/IDS-20

903 855

IDS connection cable, 25 m

AS/IDS-25

903 856

IDS connection cable, 40 m

AS/IDS-40

903 857

IDS connection cable, 60 m

AS/IDS-60

903 858

IDS connection cable, 100 m

AS/IDS-100

903 859

Содержание SensoLyt ORP 900-P

Страница 1: ...OPERATING MANUAL ba75993e03 06 2017 SensoLyt ORP 900 P IDS ORP ELECTRODE OPERATING MANUAL ...

Страница 2: ...ba75993e03 06 2017 SensoLyt ORP 900 P 2 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany ...

Страница 3: ... is used for measurement and for measured value documen tation The digital transmission technique guarantees the failure free communica tion with the meter even with long connection cables The sensor firmware can be updated via the meter Armoring A 925 P K and A 925 P S accessory To protect the electrode against mechanical impacts the A 925 P K armoring with plastic protective hood or A 925 P S wi...

Страница 4: ...ion Open the locking device step 1 Use your thumb and index finger to press the clips of the connector together and pull the connector out of the plug step 2 Closing the plug connection Make sure that the plug connection is completely dry and clean Align the guidance of the connector with the groove in the plug and insert the connector in the unlocked plug until it catches step 3 Close the locking...

Страница 5: ...e connected to a meter Accessories for the connection of the SensoLyt ORP 900 P P sensor to the meter are listed in chapter 7 WEAR PARTS AND ACCESSORIES To open and close the IDS plug in position please note the section 2 1 OPENING AND CLOSING THE PLUG CONNECTION 2 3 Calibrating measuring Conversion to standard hydrogen electrode UH UMeas URef URef depends on the reference system and temperature a...

Страница 6: ...ference electrolyte KCl 3 mol L Ag free Prior to the next measurement shortly rinse the electrode with the test sample or deionized water Overnight or longer Put the clean electrode in the watering cap that is filled with reference elec trolyte KCl 3 mol L Ag free NOTE Do not store the electrode dry or in deionized water The electrode could be permanently damaged by this T C URef mV T C URef mV Se...

Страница 7: ...S plug Synthetic materials Glass fiber reinforced Noryl TPU TPC ET POM PVC PEEK PBT O ring FPM Contacts gold plated General data ORP electrode Platinum ring Reference electrolyte Polymer electrolyte Junction 2 hole junction Shunt conduction element Ag AgCl Temperature sensor Integrated NTC 30 30 k at 25 C 77 F Measurement and application characteristics ORP measuring range 1250 0 1250 0 mV Minimum...

Страница 8: ...3 mol L Ag free Disposal Residual waste Maintenance equipment Description Model Order no Reference electrolyte solution 250 ml to fill the watering cap KCl 3 mol l Ag free KCl 250 109 705 ORP buffer solution 250 ml RH 28 109 740 Connection cable SensoLyt ORP 900 P meter Description Model Order no IDS connection cable 1 5 m AS IDS 1 5 903 850 IDS connection cable 3 m AS IDS 3 903 851 IDS connection...

Страница 9: ...r supplier Radio connection SensoLyt ORP 900 P meter Description Model Order no WLM capable IDS meter radio module for IDS meter see Internet Radio module for plug head sensor IDS WLM S 108 141 General purpose accessories Description Model Order no Blind plug for IDS plug BPO IDS 900 908 371 Armoring without protective hood A 925 P 903 838 Armoring with plastic protective hood A 925 P K 903 839 Ar...

Страница 10: ...Disposal SensoLyt ORP 900 P 10 ba75993d03 06 2017 ...

Страница 11: ......

Страница 12: ...s buildings factories and farms In more than 150 countries we have strong long standing relationships with customers who know us for our powerful combination of leading product brands and applications expertise backed by a legacy of innovation Xylem Analytics Germany GmbH Dr Karl Slevogt Str 1 82362 Weilheim Germany Xylem Analytics Germany Sales GmbH Co KG WTW Dr Karl Slevogt Str 1 82362 Weilheim ...