36
8. Analytical Quality Assurance (AQA)
AQA check OK ...
...or error message
– Repeat the check
If the error is repeated, perform troubleshooting of
the error. See "Analytical Quality Assurance" in part
"General information".
The system
AQA2 check must be performed
separately for each method monitored.
The release is stored with the date and the
specified interval. The AQA2 interval system
set up for the respective method begins again.
After approx.
2 s:
Example: Report output (AQA mode: n weeks)
AQA check system
AQA2
26.08.97
11:02
operator:
AQA interval
4 weeks
method
meas.
value
unit
theor. val.
tolerance
result
14554
1.95
mg/l
2.00
0.20
ok
Example: Report output (AQA mode: n measurements)
AQA check system
AQA2
26.08.97
11:02
operator:
AQA interval
100 measurements
method
meas.
value
unit
theor. val.
tolerance
result
14554
1.95
mg/l
2.00
0.20
ok