pH value
pH 3310 IDS
26
ba77035e01
02/2014
Continuing with two-
point calibration
Continuing with
three- to five-point
calibration
10. Thoroughly rinse the sensor with deionized water.
11. Immerse the pH sensor in buffer solution 2.
12. When measuring without temperature sensor:
Enter the temperature of the buffer with
<
><
>
.
13. Start the measurement with
<ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
14. Wait for the measurement with stability control to be completed or ter-
minate the stability control and take over the calibration value with
<ENTER>
.
The calibration display for the next buffer appears (voltage display).
15. If necessary, finish the calibration procedure as a two-point calibration
with
<M>
.
The calibration record is displayed.
pH
0.0
24.8 °C
AutoCal TEC
7.000
01.07.2011
08:00
AR
Buffer
mV
16. Thoroughly rinse the sensor with deionized water.
17. Immerse the sensor in the next buffer solution.
18. When measuring without temperature sensor:
Enter the temperature of the buffer with
<
><
>
.
19. Start the measurement with
<ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.